Based Studies were carried out to analyze internal dose for radiation worker at Boron Neutron Capture Therapy (BNCT) facility base on Cyclotron 30 MeV with BSA and room that actually design befor e. This internal dose analyze include interaction between neutron and air. The air contains N2 (72%), O2 (20%), Ar (0.93%), CO2, Neon, Kripton, Xenon, Helium and Methane. That internal dose to the worker should be bellow limit dose for radiation worker amount of 20 mSv/years. From the particle that are present in the air, only Nitrogen and Argon can change into radioactive element. Nitrogen-14 activated to Carbon-14, Nitrogen-15 activated to Nitrogen-16, and Argon-40 activated to Argon-41. Calculation using tally facility in Monte Carlo N Particle Version Extended (MCNPX) program for calculated flux Neutron in the air 3,16x10 7 Neutron/cm 2 s. room design in cancer facility have a measurement of length 200 cm, width 200 cm and high 166,40 cm. flux neutron can be used to calculated the reaction rate which is 80,1x10 -2 reaction/cm 3 s for carbon-14 and 8,75x10 -5 reaction/cm 3 s. Internal dose exposed to the radiation worker is 9.08E-9 µSv.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.