Background
To evaluate the dissociation effect of optical fluorescence platelet counting of BC-6800 hematology analyzer on ethylene diamine tetraacetic acid-dependent pseudo thrombocytopenia (EDTA-PTCP) samples.
Methods
Twenty-three finally identified EDTA-PTCP samples were recruited in this study using criteria as follow: (I) impedance platelet counts lower than 100×10
9
/L with instrument “platelet aggregation” flag; (II) existence of platelet clumps in the blood smear; (III) obviously higher platelet counts without “platelet aggregation” flag and no platelet clumps in blood smear after repeating phlebotomy using citrate anticoagulated tubes. The BC-6800 hematology analyzer and the XE-2100 hematology analyzer were used to test 23 EDTA-PTCP samples and 30 controls on both the impedance channel and the reticulocyte channel. The dissociation rate was defined as optical fluorescence platelet counts in the EDTA tubes/impedance platelet counts in citrate tubes ×100%.
Results
BC-6800 analyzer’s optical fluorescence platelet counts of EDTA-PTCP samples were significantly higher than impedance platelet counts (t=4.33, P=0.00) and comparable with the platelet counts of re-collected samples in tubes containing citrate anticoagulant. On BC-6800 hematology analyzer, 22 of 23 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 93%. On the XE-2100 hematology analyzer, 1 of the 17 EDTA-PTCP samples showed a dissociation rate greater than 80%, and the average dissociation rate was 56%.
Conclusions
Optical fluorescence platelet counting of BC-6800 Hematology Analyzer is effective for the correction of spurious low platelet counts in EDTA-PTCP patients, and its dissociation effect on EDTA-PTCP samples was independent of fluorescent dye staining.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.