overestimated. Irrespectively, it is important to point out that there is little information with regards to real-life battery longevity for the studied CRT-D device (Quadra Assura MP, St Jude) given its fairly recent commercialization. Nonetheless, we believe the main value of the present study lies in the reported proportional differences in battery longevity between the different pacing programming protocols. Our results may therefore help clinicians make more informed decisions when considering MPP activation, given the current scarcity of information regarding its impact on battery longevity.
ConclusionsIn most cases, MPP activation significantly reduces battery longevity compared with that for conventional CRT configuration. However, when reasonable MPP LV vector PCTs (< _4.0 V) are achieved, the decrease in battery longevity is relatively small and this may prompt the clinician to activate MPP in such scenarios.
Retained lead fragments following OHT occur commonly, without adverse clinical events for this cohort; however, the long-term clinical implications remain uncertain. Complete removal of all CIED hardware should be attempted at the time of OHT, and when this is not possible leads should be left in a state that facilitates their removal at a later date if required.
Aims
Cryoballoon pulmonary vein isolation (PVI) is a safe and effective treatment for atrial fibrillation (AF). Current limitations include incomplete vein occlusion due to balloon rigidity and inconsistent electrogram recording, which impairs identification of isolation. We aimed to evaluate the acute safety and performance of a novel cryoballoon system.
Methods and results
The system includes a steerable sheath, mapping catheter, and a balloon that maintains uniform inflation pressure and size following initiation of ablation. Protocol-directed cryoablation was delivered for 180 s for isolation documented in ≤60 s, otherwise freeze duration was 240 s. Primary endpoints were acute safety and vein isolation. Pulmonary vein isolation was confirmed at ≥30 min post-isolation. Data were compared across vein locations. Thirty patients with paroxysmal AF were enrolled at two centres and underwent PVI. Pulmonary vein isolation was achieved with cryoablation only in 100% of veins (120/120). Nadir temperature was −53.1 ± 5.3°C. The number of applications to achieve PVI was 1.4 ± 0.4 per vein. Of the 120 veins, 89 were isolated with a single cryothermal application (10/30 patients required only 4 total cryoablations). There were no procedural- or device-related serious adverse events at 30 days post-procedure. A subset (24/30) of patients was followed for 1-year and 71% (17/24) remained free of atrial arrhythmias. Six patients with arrhythmia recurrence were remapped and three had durable PVI for all four veins.
Conclusion
In this first human experience, the novel cryoballoon platform was safe, efficacious, and demonstrated a high proportion of successful single ablation isolation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.