A computational model has been developed to quantify the degree of cross-shear of a polyethylene pin articulating against a metallic plate, based on the direct simulation of a multidirectional pin-onplate wear machine. The principal molecular orientation (PMO) was determined for each polymer site. The frictional work in the direction perpendicular to the PMO was assumed to produce the greatest orientation softening (Wang et al., 1997). The cross-shear ratio (CS) was defined as the frictional work perpendicular to the PMO direction, divided by the total frictional work. Cross-shear on the pin contact surface was location-specific, and of continuously changing magnitude because the direction of frictional force continuously changed due to pin rotation. The polymer pin motion was varied from a purely linear track (CS=0) up to a maximum rotation of ±55° (CS=0.254). The relationship between wear factors (K) measured experimentally and theoretically predicted CS was defined using logarithmic functions for both conventional and highly cross-linked UHMWPE. Crossshear increased the apparent wear factor for both polyethylenes by more than 5-fold compared to unidirectional wear.
Cross-linked polyethylene currently is being introduced in knee prostheses. The wear rates, wear debris, and biologic reactivity of non cross-linked, moderately cross-linked, and highly cross-linked polyethylene have been compared in multidirectional wear tests and knee simulators. Multidirectional pin-on-plate wear studies of noncross-linked, moderately cross-linked (5 Mrad), and highly cross-linked (10 Mrad) polyethylene showed a 75% reduction in wear with the highly cross-linked material under kinematics found in the hip, but only a 33% reduction under wear in kinematics representative of the knee. In knee simulator studies, with the fixed-bearing press-fit, condylar Sigma cruciateretaining knee under high kinematic input conditions, the wear of 5 Mrad moderately cross-linked polyethylene was 13 ± 4 mm 3 per 1 million cycles, which was lower (p < 0.05) than the wear of clinically used, gamma vacuum foil GUR 1020 polyethylene (23 ± 6 mm 3 /1 million cycles). For the lowcontact stress mobile-bearing knee, the wear of moderately cross-linked polyethylene was 2 ± 1 mm 3 per 1 million cycles, which was lower (p < 0.05) than GVF GUR 1020 polyethylene (5 ± 2 mm 3 /1 million cycles). The wear debris isolated from the fixed-bearing knees showed the moderately crosslinked material had a larger percentage volume of particles smaller than 1 µm in size, compared with GVF GUR 1020 polyethylene. Direct cell culture studies of wear debris generated in sterile wear simulators using multidirectional motion showed a increase (p < 0.05) in tumor necrosis factoralpha levels and reactivity for GUR 1050 cross-linked polyethylene debris compared with an equivalent volume of noncross-linked GUR 1050 polyethylene. The use of crosslinked polyethylene in the knee reduces the volumetric wear rate. However, the clinical significance of reduced fracture toughness, elevated wear in abrasive conditions, and the elevated tumor necrosis factor-alpha release from smaller more reactive particles warrant further investigation.Intentionally cross-linked polyethylene (PE) has been introduced extensively into hip prostheses during the last 5 years. Initial laboratory studies in hip simulators indicated that with high levels of cross-linking produced by 10 Mrad of irradiation, virtually 0 wear was produced. 2,13,14 More recently, both clinical and simulator studies of highly cross-linked PE done in more physiologically relevant conditions 5 showed a reduction (p < 0.05) in wear compared with conventional materials, with finite wear rates in the range of 5 to 10 mm 3 per 1 million cycles. Investigation of the wear debris from hip simulator studies has shown that moderately cross-linked PE has a larger percentage number and volume of smaller particles, compared with PE that has not been cross-linked. 4 These differences can only been shown in studies that analyze the whole size range of PE debris, and this is shown more clearly when the volume of wear debris rather than number of particles is analyzed. These observations are particularly important, as i...
Artificial joints employing ultra-high molecular weight polyethylene (UHMWPE) are widely used to treat joint diseases and trauma. Wear of the polymer bearing surface largely limits the use of these joints in younger and more active patients. Previous studies have shown the wear factor used in Archard's law for the conventional polyethylene to be highly dependent on contact pressure and this has produced variability in experimental data and has constrained the reliability and applicability of previous computational predictions. A new wear law is proposed, based on wear volume being dependent on, and proportional to, the product of the sliding distance and contact area. The dimensionless proportional constant, wear coefficient, which was independent of contact pressure, was determined from a multi-directional pin on plate study. This was used in computational predictions of the wear of the conventional UHMWPE hip joints. The wear of the polyethylene cup was independently experimentally determined in physiological full hip joint simulator studies. The predicted wear rate from the new computational model was generally increased, with an improved agreement with the experimental measurement compared with the previous computational model. It was shown that wear in the UHMWPE hip joints increased as head size and contact area increased. This resulted in a much larger increase in the wear rate as the head size increased, compared with the previous computational model, and is consistent with clinical observations. This new understanding of the wear mechanism in artificial joints using the UHMWPE bearing surfaces, and the improved ability to predict wear independently and to address previously described discrepancies offer new opportunities to optimize design parameters.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.