While locking plate fixation is becoming increasingly popular for complex and osteoporotic fractures, for many indications compression plating remains the standard choice. This study compares the mechanical behaviour of the more recent locking compression plate (LCP) device, with the traditional dynamic compression plates (DCPs) in bone of varying quality using finite element modelling. The bone properties considered include orthotropy, inhomogeneity, cortical thinning and periosteal apposition associated with osteoporosis. The effect of preloads induced by compression plating was included in the models. Two different fracture scenarios were modelled: one with complete reduction and one with a fracture gap. The results show that the preload arising in DCPs results in large principal strains in the bone all around the perimeter of the screw hole, whereas for LCPs large principal strains occur primarily on the side of the screw proximal to the load. The strains within the bone produced by the two screw types are similar in healthy bone with a reduced fracture gap; however, the DCP produces much larger strains in osteoporotic bone. In the presence of a fracture gap, the DCP results in a considerably larger region with high tensile strains and a slightly smaller region with high compressive strains. These findings provide a biomechanical basis for the reported improved performance of locking plates in poorer bone quality.
When using locked plating for bone fracture fixation, screw loosening is reported as one of the most frequent complications and is commonly attributed to an incorrect choice of screw configuration. Choosing a patient-optimized screw configuration is not straightforward as there are many interdependent variables that affect device performance. The aim of the study was to evaluate the influence that locking screw configuration has on loosening risk and how this is influenced by bone quality. This study uses finite element models that incorporate cortical bone heterogeneity, orthotropy, and geometrical nonlinearity to examine the effect of screw configuration on variables associated with loosening and interfragmentary motion. Strain levels within the bone were used as indicators of regions that may undergo loosening. The study found that, in healthy bone under axial loading, the most important variables influencing strain levels within the bone were the size of the bridging span (working length) and the plate rigidity. Unlike healthy bone, osteoporotic bone was found to be particularly sensitive to the spacing of the screws within the plate. Using two empty screw holes between the screws closest to the fracture was found to reduce the strain levels at the first screw by 49% in osteoporotic bone (compared to only 2.4% in healthy bone). The study also found that under torsional loading the total number of screws used was the most important variable with a 59% reduction in the strain around the screws closest to the fracture when using six rather than four screws in osteoporotic bone. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:1856-1864, 2016.
ObjectivesOpening wedge high tibial osteotomy (HTO) is an established surgical procedure for the treatment of early-stage knee arthritis. Other than infection, the majority of complications are related to mechanical factors – in particular, stimulation of healing at the osteotomy site. This study used finite element (FE) analysis to investigate the effect of plate design and bridging span on interfragmentary movement (IFM) and the influence of fracture healing on plate stress and potential failure.Materials and MethodsA 10° opening wedge HTO was created in a composite tibia. Imaging and strain gauge data were used to create and validate FE models. Models of an intact tibia and a tibia implanted with a custom HTO plate using two different bridging spans were validated against experimental data. Physiological muscle forces and different stages of osteotomy gap healing simulating up to six weeks postoperatively were then incorporated. Predictions of plate stress and IFM for the custom plate were compared against predictions for an industry standard plate (TomoFix).ResultsFor both plate types, long spans increased IFM but did not substantially alter peak plate stress. The custom plate increased axial and shear IFM values by up to 24% and 47%, respectively, compared with the TomoFix. In all cases, a callus stiffness of 528 MPa was required to reduce plate stress below the fatigue strength of titanium alloy.ConclusionWe demonstrate that larger bridging spans in opening wedge HTO increase IFM without substantially increasing plate stress. The results indicate, however, that callus healing is required to prevent fatigue failure.Cite this article: A. R. MacLeod, G. Serrancoli, B. J. Fregly, A. D. Toms, H. S. Gill. The effect of plate design, bridging span, and fracture healing on the performance of high tibial osteotomy plates: An experimental and finite element study. Bone Joint Res 2018;7:639–649. DOI: 10.1302/2046-3758.712.BJR-2018-0035.R1.
ObjectivesSecondary fracture healing is strongly influenced by the stiffness of the bone-fixator system. Biomechanical tests are extensively used to investigate stiffness and strength of fixation devices. The stiffness values reported in the literature for locked plating, however, vary by three orders of magnitude. The aim of this study was to examine the influence that the method of restraint and load application has on the stiffness produced, the strain distribution within the bone, and the stresses in the implant for locking plate constructs.MethodsSynthetic composite bones were used to evaluate experimentally the influence of four different methods of loading and restraining specimens, all used in recent previous studies. Two plate types and three screw arrangements were also evaluated for each loading scenario. Computational models were also developed and validated using the experimental tests.ResultsThe method of loading was found to affect the gap stiffness strongly (by up to six times) but also the magnitude of the plate stress and the location and magnitude of strains at the bone-screw interface.ConclusionsThis study demonstrates that the method of loading is responsible for much of the difference in reported stiffness values in the literature. It also shows that previous contradictory findings, such as the influence of working length and very large differences in failure loads, can be readily explained by the choice of loading condition.Cite this article: A. MacLeod, A. H. R. W. Simpson, P. Pankaj. Experimental and numerical investigation into the influence of loading conditions in biomechanical testing of locking plate fracture fixation devices. Bone Joint Res 2018;7:111–120. DOI: 10.1302/2046-3758.71.BJR-2017-0074.R2.
Background Millions of non-locking screws are manually tightened during surgery each year, but their insertion frequently results in overtightening and damage to the surrounding bone. We postulated that by calculating the torque limit of a screw hole, using bone and screw properties, the risk of overtightening during screw insertion could be reduced. Additionally, predicted maximum torque could be used to identify optimum screw torque, as a percentage of the maximum, based on applied compression and residual pullout strength. Methods Longitudinal cross-sections were taken from juvenile bovine tibial diaphyses, a validated surrogate of human bone, and 3.5 mm cortical non-locking screws were inserted. Fifty-four samples were used to define the association between stripping torque and cortical thickness. The relationship derived enabled prediction of insertion torques representing 40 to 100% of the theoretical stripping torque (T str) for a further 170 samples. Screw-bone compression generated during insertion was measured, followed immediately by axial pullout testing. Findings Screw-bone compression increased linearly with applied torque up to 80% of T str (R 2 =0.752, p<0.001), but beyond this, no significant further compression was generated. After screw insertion, with all screw threads engaged, more tightening did not create any significant (R 2 =0.000, p=0.498) increase in pullout strength. Interpretation Increasing screw tightness beyond 80% of the maximum did not increase screw-bone compression. Variations in torques below T str , did not affect pullout forces of inserted screws. Further validation of these findings in human bone and creation of clinical guidelines based on this research approach should improve surgical outcomes and reduce operative costs.
ObjectivesModular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection.Materials and MethodsTi-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Results36 mm diameter heads had significantly lower pull-off forces than 28 mm heads when impacted at 4 kN and 5 kN (p < 0.001; p < 0.001), but not at 6 kN (p = 0.21). Mean pull-off forces at 4 kN and 5 kN impaction forces were approximately 20% larger for 28 mm heads compared with 36 mm heads. Finite element and analytical models demonstrate that the differences in pull-off strength can be explained by differences in structural rigidity and the resulting interface pressures.ConclusionThis is the first study to show that 36 mm Co-Cr heads have up to 20% lower pull-off connection strength compared with 28 mm heads for equivalent assembly forces. This effect is likely to play a role in the high failure rates of large diameter MoM hips.Cite this article: A. R. MacLeod, N. P. T. Sullivan, M. R. Whitehouse, H. S. Gill. Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability. Bone Joint Res 2016;5:338–346. DOI: 10.1302/2046-3758.58.BJR-2016-0044.R1.
Most locked plating failures are due to inappropriate device configuration for the fracture pattern. Several studies cite screw positioning variables such as the number and spacing of screws as responsible for occurrences of locking plate breakage, screw loosening, and peri-prosthetic re-fracture. It is also widely accepted that inappropriate device stiffness can inhibit or delay healing. Careful preoperative planning is therefore critical if these failures are to be prevented. This study examines several variables which need to be considered when optimising a locking plate fixation device for fracture treatment including: material selection; screw placement; the effect of the fracture pattern; and the bone-plate offset. We demonstrate that device selection is not straight-forward as many of the variables influence one-another and an identically configured device can perform very differently depending upon the fracture pattern. Finally, we summarise the influence of some of the key parameters and the influence this can have on the fracture healing environment and the stresses within the plate in a flowchart.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.