Current methods for high resolution imaging of neural activity, such as functional magnetic resonance imaging and penetrating electrode arrays, are not well suited for chronic patient monitoring in clinical settings. The ability to locally monitor neural activity in the brain would benefit doctors and surgeons treating and diagnosing neurological diseases such as epilepsy. Measurement of the intrinsic optical signal offers a high resolution (better than 100 microns) for the localization of brain activity. The localized detection of seizures in rats using an implantable optical sensor is demonstrated. The implantable optical sensor is fabricated on a 25 µm thick, flexible polyimide substrate and uses microLEDs to measure localized changes in reflectivity due to cerebral blood flow when attached to the surface of the brain. Quantum dot coatings are applied directly to the microLEDs via pneumatic printing for wavelength conversion by photoluminescent re-emission.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.