An accurate, selective and sensitive spectrofluorimetric method was developed for the determination of citalopram hydrobromide in commercial dosage forms. The method was based on the formation of a fluorescent ion-pair complex between citalopram hydrobromide and eosin Y in the presence of a disodium hydrogen phosphate/citric acid buffer solution of pH 3.4 that was extractable in dichloromethane. The extracted complex showed fluorescence intensity at λem = 554 nm after excitation at 259 nm. The calibration curve was linear over at concentrations of 2.0-26.0 µg/mL. Under optimized experimental conditions, the proposed method was validated as per ICH guidelines. The effect of common excipients used as additives was tested and the tolerance limit calculated. The limit of detection for the proposed method was 0.121 μg/mL. The proposed method was successfully applied to the determination of citalopram hydrobromide in commercial dosage forms. The results were compared with the reference RP-HPLC method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.