U‐Pb Zircon Geochronology with an Integrated LA‐ICP‐MS Microanalytical Workstation: Achievements in Precision and Accuracy

Abstract: Laser ablation‐inductively coupled plasma‐mass spectrometry (LA‐ICP‐MS) is a microanalytical tool especially suitable for providing fast and precise U‐Pb geochronological results on zircon grains. A new 193 nm excimer laser adapted to a micromachining workstation, equipped with a newly designed two‐volume ablation cell and coupled with a quadrupole ICP‐MS, is presented here. The system was tuned routinely to achieve sensitivities in the range of 3000 cps/μg g−1 for 238U (< 2% RSD), with a 34 μm spot size, at 5… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

2
63
0
33

Year Published

2015
2015
2019
2019

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 190 publications
(98 citation statements)
references
References 49 publications
2
63
0
33
Order By: Relevance
“…The analytical work was undertaken by using a "Resonetics Resolution M50" 193 nm laser "Excimer of ArF" connected to a "Thermo Xii Series Quadrupole Mass Spectrometer" following analytical procedures and technical details after Solari et al (2010). Selected zircon minerals were mounted in epoxy and polished.…”
Section: Appendixmentioning
confidence: 99%
See 1 more Smart Citation
“…The analytical work was undertaken by using a "Resonetics Resolution M50" 193 nm laser "Excimer of ArF" connected to a "Thermo Xii Series Quadrupole Mass Spectrometer" following analytical procedures and technical details after Solari et al (2010). Selected zircon minerals were mounted in epoxy and polished.…”
Section: Appendixmentioning
confidence: 99%
“…Analyzed spots were 23 micrometers, using an analytical protocol modified from Solari et al (2010), Geostandards Geoanalyt Res.…”
Section: Cs2-4 Spotsmentioning
confidence: 99%
“…De la muestra XO-1204 se analizaron en total 20 cristales de circón; mientras que de la muestra XO-1205 se analizaron en total 18 cristales y por último, de la muestra XO-1210 se analizaron 26 cristales (Tabla 3). Los cristales de circón se montaron en portaobjetos de vidrio y se pulieron para ser analizados por espectrómetro de masas con acoplamiento inductivo con ablación láser (LA-ICP-MS, por sus siglas en inglés) siguiendo la metodología descrita en Solari et al (2010) para geocronología por U-Pb. Estos análisis se llevaron a cabo en el Laboratorio de Estudios Isotópicos del Centro de Geociencias, UNAM, Juriquilla, Querétaro, México.…”
Section: Geocronología U-pb En Circonesunclassified
“…The signal reaches 95% of peak intensity in ~1 sec after the laser is fired (Figure 1). Similarly, the washout time is also reduced to <2 sec to recede to <5% of peak value as against 7-12 sec in case of HelEx-I design 23 or other two-volume 24 or single, smallvolume sample cell designs 25 . Such fast signal response and washout time in HelEx-II can reduce between sample washout times, thereby significantly improving sample throughput.…”
Section: Performance Of Two-volume Helex-ii Sample Ablation Chambermentioning
confidence: 99%