“…In another study, a dosimetry system comprised of the Model 7200 Automatic TLD Reader (RSD Co., Tehran, Iran) and the TLD-100 (LiF:Mg,Ti) dosimeters (Thermo Fisher Scientific) was examined for non-linearity of the response, overload, post-irradiation effects, and reusability (radiation characteristics); reader stability and power supply (electromagnetic characteristics); and light exposure and ambient temperature (environmental characteristics) (Mohammadi and Pooya 2017). Similar research of TL dosimetry systems’ properties included type testing of LiF based TLD (Radcard, Kraków, Poland) read on a RADOS RE2000 reader (Mirion Technology, Inc., Atlanta, GA) and Manual Reader-Analyzer TL RA’94 (MICROLAB, Warsaw, Poland) (Obryk et al 2008, 2011).…”