2022
DOI: 10.1515/cclm-2022-0177
|View full text |Cite
|
Sign up to set email alerts
|

Two-site evaluation of the Roche Elecsys Vitamin D total III assay

Abstract: Objectives The high request for vitamin D testing in the last decades has led manufacturers to develop assays on automated immunoassay platforms. The objective of this study was to evaluate the performance of the new Elecsys Vitamin D total III assay for the measurement of total 25(OH)D. Methods A total of 844 serum samples collected in two clinical laboratories were used to evaluate the new Roche Elecsys Vitamin D total III … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
0
0
2

Year Published

2022
2022
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(2 citation statements)
references
References 24 publications
0
0
0
2
Order By: Relevance
“…Weiterhin sind Methoden, die mit biotinilierten Reagenzien arbeiten, anfällig für eine Beeinflussung durch von Patienten eingenommene Biotinhaltige Nahrungsergänzungsmittel (teilweise werden für die Stimulation des Haarwuchses hohe Dosen verwendet) 10 .…”
Section: Introductionunclassified
See 1 more Smart Citation
“…Weiterhin sind Methoden, die mit biotinilierten Reagenzien arbeiten, anfällig für eine Beeinflussung durch von Patienten eingenommene Biotinhaltige Nahrungsergänzungsmittel (teilweise werden für die Stimulation des Haarwuchses hohe Dosen verwendet) 10 .…”
Section: Introductionunclassified
“…Spezifischer erfolgt die Messung mittels LC-MS/MS-Methoden, die keine Kreuzreaktion mit 24,25-dihydroxyvitamin aufweisen. Jedoch kann bei den LC-MS/MS-Methoden das 3-epi-25-OH-Vitamin D die Messung stören aufgrund des identischen Verhältnisses von Masse zu Ladung 10 13 . Wegen des erhöhten Aufwandes werden LC-MS/MS -Methoden in der Routine -Diagnostik nicht häufig eingesetzt.…”
Section: Introductionunclassified