Abstract: Purpose The RefleXion X1 is a novel radiotherapy machine designed for image‐guided radiotherapy (IGRT) and biology‐guided radiotherapy (BgRT). Its treatment planning system (TPS) generates IMRT and SBRT plans for a 6MV‐FFF beam delivered axially via 50 firing positions with the couch advancing every 2.1 mm. The purpose of this work is to report the TPS commissioning results for the first clinical installation of RefleXion™ X1. Methods CT images of multiple phantoms were imported into the RefleXion TPS to evalu… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
8
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
7

Relationship

3
4

Authors

Journals

citations
Cited by 12 publications
(8 citation statements)
references
References 28 publications
0
8
0
Order By: Relevance
“…Simiele et al reported on the results of the commissioning of the X1 TPS. 8 The results showed that the TPS met the specified tolerances for dose calculation accuracy and data transfer for targets >1.5 cm in diameter. The end-to-end testing showed small targeting errors for both isocentric and off-axis treatments.…”
Section: Discussionmentioning
confidence: 89%
See 1 more Smart Citation
“…Simiele et al reported on the results of the commissioning of the X1 TPS. 8 The results showed that the TPS met the specified tolerances for dose calculation accuracy and data transfer for targets >1.5 cm in diameter. The end-to-end testing showed small targeting errors for both isocentric and off-axis treatments.…”
Section: Discussionmentioning
confidence: 89%
“…Our previous phantom study found the presence of an interplay effect when treating targets moving along the direction of table motion. 8 To prevent the interplay effect, only tumors with an amplitude of motion of <2 cm are treated based on the motion-inclusive target volume. During treatment, a slow couch speed CT scan is acquired to maintain the target motion information and enable accurate patient positioning.…”
Section: Discussionmentioning
confidence: 99%
“…The quality assurance for all tested plans passed gamma 3%/2 mm criteria as described in a related X1 TPS commissioning manuscript. 6…”
Section: Reflexion Treatment Planningmentioning
confidence: 99%
“…The first clinical RefleXion X1 system was recently installed at our institution, commissioned [6][7][8] , and is currently in clinical use for IMRT and SBRT since May 2021. This work is the first treatment planning study investigating the capabilities of the clinical X1 TPS to determine the achievable plan quality for various common treatment sites including conventionally fractionated lung, head-and-neck, prostate, anus, brain, and lung SBRT.…”
Section: Introductionmentioning
confidence: 99%
“…A recently designed radiotherapy platform has incorporated an on-board kV fan-beam CT (kVCT) and Positron Emission Tomography (PET) system into a ring-shaped gantry that can provide both anatomical and functional images [8] . A few studies on this novel radiotherapy platform have been reported regarding the initial machine commissioning [9] , the characteristics of the PET imaging system performance [10] , treatment planning experiences [11] , [12] as well as a detailed clinical workflow for using this system for biology-guided radiotherapy [13] . However, the imaging performance of the kVCT system has not been investigated.…”
Section: Introductionmentioning