The use of log file analysis within VMAT audits

Abstract: Objective: This work investigated the delivery accuracy of different Varian linear accelerator models using log file-derived multileaf collimator (MLC) root mean square (RMS) values. Methods: Seven centres independently created a plan on the same virtual phantom using their own planning system, and the log files were analyzed following delivery of the plan in each centre to assess MLC positioning accuracy. A single standard plan was also delivered by the seven centres to remove variations in complexity, and th… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
27
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 22 publications
(29 citation statements)
references
References 19 publications
2
27
0
Order By: Relevance
“…Figure 5 shows the daily distributions of the maxRMS by DFV for 5 different pcsr QA plans and 12 real treatment plans which were grouped into the corresponding plan class. Similar MLC position error was revealed by different other previous works [10][11][12][13].…”
Section: Mlc Movement Analysissupporting
confidence: 87%
“…Figure 5 shows the daily distributions of the maxRMS by DFV for 5 different pcsr QA plans and 12 real treatment plans which were grouped into the corresponding plan class. Similar MLC position error was revealed by different other previous works [10][11][12][13].…”
Section: Mlc Movement Analysissupporting
confidence: 87%
“…Magnitude of MLC errors (RMS <0.3 mm) and maximum gantry angle deviation (<0.2 degree) in this study for VMAT SBRT were 13,14 Minor differences between studies might be due to the differences of MLC positions, gantry angles, and differential dose indices with magnitudes of 1 mm, 1 degree, and 5 % respectively, while the passing rate of gamma analysis from ArcCheck TM measurement was still higher than 95 % tolerance that would be recognized as a pass…”
Section: Discussionmentioning
confidence: 51%
“…12 Several studies have been performed to evaluate the efficiency and effectiveness of machine log file analysis for volumetric-modulated arc therapy (VMAT). 13,14 Chan et al conducted a preliminary study on quantifying the deviations of monitor unit (MU) and gantry angles during delivery against the plan through log analysis for three head and neck and three prostate VMAT cases. 14 Mcgarry et al quantified the plan delivery accuracy in terms of log file-derived MLC positioning error with single center-specific plan and single standard plan for each participating center in VMAT audits.…”