The PTB Beta Secondary Standard BSS 2 for radiation protection

Abstract: With the Beta Secondary Standard (BSS), an important basis was created thirty years ago for the dissemination of the unit of the absorbed dose rate of beta radiation to tissue, which has proven its worth in practice. The main reasons for the popularity of the BSS are the traceability of its calibration to the primary standard measuring device of PTB and its conformity with the recommendations of the international standard ISO 6980. In 1996, an extensive modification of the BSS was started with the aim of impro… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
39
0
1

Year Published

2012
2012
2019
2019

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 46 publications
(43 citation statements)
references
References 3 publications
0
39
0
1
Order By: Relevance
“…The filter assures that the radiation field is more homogeneous in the irradiation plane than without it. It is located at a distance of 10 cm from the radiation source [7]. In the last two lines of table 1 the position of the phase space file planes (location where the simulation starts) is given.…”
Section: Methods Of Simulationmentioning
confidence: 99%
“…The filter assures that the radiation field is more homogeneous in the irradiation plane than without it. It is located at a distance of 10 cm from the radiation source [7]. In the last two lines of table 1 the position of the phase space file planes (location where the simulation starts) is given.…”
Section: Methods Of Simulationmentioning
confidence: 99%
“…Thin layer TLDs [4] were used to measure in mixed beta-gamma radiation fields, while for the X-Ray CT-measurements "Photon-TLDs" of type MTS-N [5] were employed. The "Beta-TLDs" were calibrated with a Beta Secondary Standard 2 [6], while the Photon-TLDs were calibrated according to…”
Section: Methodsmentioning
confidence: 99%
“…Two different types of beta-ray irradiation systems have been installed and served as beta reference radiation elds at the Facility of Radiation Standards (FRS) in the Japan Atomic Energy Agency (JAEA). One is so-called the Beta Secondary Standard type 2 (BSS2), 1) which is commercially available and commonly used worldwide as a de facto standard in beta calibration. The BSS2 was newly installed in 2010 at the JAEA-FRS.…”
Section: Original Paper I I Troductiomentioning
confidence: 99%
“…For a dosemeter composed of low-Z meterials, the same ,Sr + Y sys (7 mg/cm 2 ) (mGy/h) * 1 79.4 ± 0.9 * 2 34.5 ± 0.7 * 2 * 1 These values were obtained directly from the measurements using extrapolation chamber. * 2 As of 13 th November 2013 and the associated uncertainties are given for k = 1. equation can be applied by replacing the values u and d 1 with their tissue-equivalent ones which is obtained by multiplying the scaling factor relative to the tissue given in ISO 6980-2.…”
Section: Depth-dose Curvesmentioning