Tests of SNIS Josephson Arrays Cryocooler Operation

Abstract: Cryogen-free operation of is essential to spread applications of superconductivity and is indeed unavoidable in some cases. In electrical metrology applications, higher temperature operation to reduce the refrigerator size and complexity is not yet possible, since arrays of Josephson junctions for voltage standard applications made with high-temperature superconductors are not yet available. The superconductor-normal metal-insulator-superconductor (SNIS) technology developed at INRIM uses low temperature super… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
0

Year Published

2015
2015
2021
2021

Publication Types

Select...
5
2

Relationship

2
5

Authors

Journals

citations
Cited by 7 publications
(4 citation statements)
references
References 15 publications
0
4
0
Order By: Relevance
“…When comparing helium gas with cryocooler data [17], one can see that the step is reduced in the latter case from approximately 0.5 mA to about 0.2 mA. In the inset a higher resolution plot of the step profile is shown for a lower order Fig.…”
Section: B Rf Step Propertiesmentioning
confidence: 90%
“…When comparing helium gas with cryocooler data [17], one can see that the step is reduced in the latter case from approximately 0.5 mA to about 0.2 mA. In the inset a higher resolution plot of the step profile is shown for a lower order Fig.…”
Section: B Rf Step Propertiesmentioning
confidence: 90%
“…The Si diodes temperature was read with a cryogenic controller, a Lake Shore 350 (LS350, henceforth). Both Cu carrier and coldplate are thermally and electromagnetically shielded in the setup realized for high accuracy measurements for metrology [16]. A manganin wire of resistance R h ≃ 80 Ω at 4.2 K is wound around the coldplate disk as heater, when the temperature is closed-loop controlled.…”
Section: Josephson Chip and Cryogenic Equipmentmentioning
confidence: 99%
“…The cold finger top was fitted with an additional disk retaining inside a deep-set thermometer that, along with a heater wire wound around it, allows fine temperature monitoring and control. A second thermometer close to the chip was used to better estimate the array operating conditions and detect unwanted thermal gradients [15].…”
Section: Experimental and Cryocooler Set-upmentioning