2015
DOI: 10.1118/1.4930058
|View full text |Cite
|
Sign up to set email alerts
|

Technical Note: Initial characterization of the new EBT‐XD Gafchromic film

Abstract: The LRA of EBT-XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

2
41
0

Year Published

2016
2016
2023
2023

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 35 publications
(46 citation statements)
references
References 31 publications
2
41
0
Order By: Relevance
“…Guerda Massillon et al (8) reported that the response of EBT3 films only weakly depends on the energy of high‐energy photon beams (6, 15 MV) that are generally used in radiotherapy, and demonstrated a variation under 11% for very low‐energy photons (50 kV). Grams et al (17) demonstrated minimal energy dependence for EBT‐XD films irradiated with 6–18 MV beams (<0.8%). Further studies are needed to determine the characteristics of EBT‐XD films exposed to low‐energy photon, electron, and proton beams, over a broad range of energies.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Guerda Massillon et al (8) reported that the response of EBT3 films only weakly depends on the energy of high‐energy photon beams (6, 15 MV) that are generally used in radiotherapy, and demonstrated a variation under 11% for very low‐energy photons (50 kV). Grams et al (17) demonstrated minimal energy dependence for EBT‐XD films irradiated with 6–18 MV beams (<0.8%). Further studies are needed to determine the characteristics of EBT‐XD films exposed to low‐energy photon, electron, and proton beams, over a broad range of energies.…”
Section: Discussionmentioning
confidence: 99%
“…Before performing patient‐specific QA of radiotherapy treatment plans with the Gafchromic EBT‐XD films, it is important to investigate their dosimetric characteristics. It has been reported that the lateral response artifacts associated with EBT‐XD films are significantly smaller than those associated with EBT3 films 17 , 18 …”
Section: Introductionmentioning
confidence: 99%
“…The active layer of the EBT‐XD model is sandwiched between two matte polyester substrate each with 125 μm thickness, identical to that of the EBT3 model. There are several studies on the characterization and validation of the EBT‐XD films for radiotherapy dosimetry applications . These studies showed an improved performance of the EBT‐XD film over the EBT3 model for applications requiring doses greater than 10 Gy.…”
Section: Introductionmentioning
confidence: 99%
“…Extended range Gafchromic film (EBT‐XD, Ashland Inc., Bridgewater, NJ) was chosen because of the wide dynamic range,26, 27 well‐suited for SRS verification. An additional benefit is greatly reduced scanner lateral response artifact 26.…”
Section: Methodsmentioning
confidence: 99%
“…An additional benefit is greatly reduced scanner lateral response artifact 26. Both calibration and measurement film pieces were scanned in the same orientation with respect to the original sheet they came from.…”
Section: Methodsmentioning
confidence: 99%