2014
DOI: 10.1118/1.4898598
|View full text |Cite
|
Sign up to set email alerts
|

Technical Note: Influence of the phantom material on the absorbed‐dose energy dependence of the EBT3 radiochromic film for photons in the energy range 3 keV–18 MeV

Abstract: The EBT3 film shows a reasonably constant absorbed-dose energy dependence when irradiated in water. If the dose-to-water in the phantom is considered, the maximum difference of EBT3 film energy dependence with the solid phantoms studied with respect to water is about 6% (at an energy of 50 keV). The reported overall energy dependence of the EBT3 film in water at energies below 100 keV is mainly due to the intrinsic energy dependence.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
24
0
4

Year Published

2015
2015
2020
2020

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 19 publications
(30 citation statements)
references
References 19 publications
2
24
0
4
Order By: Relevance
“…The intrinsic energy response may be written:knormalbq=Mnormalfilmfalse(Qfalse)Dnormalfilmfalse(Qfalse),where a radiation beam with quality Q produces an absorbed dose in film, D film ( Q ), and M film ( Q ) is the film response measurement (note that this quantity is defined inversely from Sutherland and Rogers, Bekerat et al ., and Hermida‐López et al …”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…The intrinsic energy response may be written:knormalbq=Mnormalfilmfalse(Qfalse)Dnormalfilmfalse(Qfalse),where a radiation beam with quality Q produces an absorbed dose in film, D film ( Q ), and M film ( Q ) is the film response measurement (note that this quantity is defined inversely from Sutherland and Rogers, Bekerat et al ., and Hermida‐López et al …”
Section: Methodsmentioning
confidence: 99%
“…found contrasting magnitudes of the total energy dependence of EBT3 film to low‐energy photons of up to 11% and 3%, respectively. This inconsistency has been attributed to manufacturing and compositional changes that are not always accompanied by updates to the film model or label . Using Monte Carlo simulations of two batches of EBT2 film with slight variations in active layer chemical composition, Sutherland and Rogers showed that the magnitude of energy‐dependent response for photons under 100 keV is highly dependent on the particular batch makeup (one batch had 10% and the other a 50% energy response).…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Similarly, simulations by Hermida-Lopez et al [57] indicated that the EBT and EBT2 films exhibit an energy-dependent response in water in the energy region from 10 to 100 keV, with a 10% and 40% maximum reduction at 40 keV, respectively. They predicted that the EBT3 films would have a constant response within 2.3% over the entire energy region.…”
Section: Radiochromic Film Dosimetrymentioning
confidence: 89%
“…16 The most recent EBT3 model also introduced changes to the active layer composition to improve the energy dependence in a wide energy range. 18,19 This work presents a practical method for measuring in water, planar absorbed dose distributions and the reference dose rates produced by 106 Ru/ 106 Rh ophthalmic plaques using EBT3 radiochromic film. The measurement results are compared with published simulation and experimental data.…”
Section: Introductionmentioning