1984
DOI: 10.1016/s0378-4347(00)83349-6
|View full text |Cite
|
Sign up to set email alerts
|

Semi-automated high-performance liquid chromatographic method for the determination of cyclosporine in plasma and blood using column switching

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

1984
1984
2006
2006

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 42 publications
(1 citation statement)
references
References 13 publications
0
1
0
Order By: Relevance
“…method of Smith & Robinson (1984) wZs adapted as follows: to 0.5 ml whole blood, 0.1 ml methanol was added and vortexed for 5 s. Then 1.2 ml of internal standard solution (250 ng/ml of Cyclosporin D in acetonitrile/water 97.5/2.5 v/v) were added, the tubes vortexed again for 30 s and then centrifuged for 10 min at 1,400 rev/min. The clear supernatant Ir was decanted into a conical glass tube to which 1 ml N-hexane was added.…”
Section: Protocolmentioning
confidence: 99%
“…method of Smith & Robinson (1984) wZs adapted as follows: to 0.5 ml whole blood, 0.1 ml methanol was added and vortexed for 5 s. Then 1.2 ml of internal standard solution (250 ng/ml of Cyclosporin D in acetonitrile/water 97.5/2.5 v/v) were added, the tubes vortexed again for 30 s and then centrifuged for 10 min at 1,400 rev/min. The clear supernatant Ir was decanted into a conical glass tube to which 1 ml N-hexane was added.…”
Section: Protocolmentioning
confidence: 99%