2017
DOI: 10.1111/pace.13159
|View full text |Cite
|
Sign up to set email alerts
|

Safety and effectiveness of a 6‐French MRI conditional pacemaker lead: The INGEVITYTM clinical investigation study results

Abstract: The clinical performance of the INGEVITY lead demonstrated a high lead-related complication-free rate over 12 months of follow-up and excellent electrical characteristics.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
4
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(4 citation statements)
references
References 24 publications
0
4
0
Order By: Relevance
“…These leads incorporate design aspects for MRI conditional safety. The inner coil of the MRI lead was modified to have a higher inductance to prevent heating during MRI scanning, which was achieved through the unifilar design of the inner coils of the INGEVITY leads [ 13 , 15 , 16 ]. However, because torque transfer decreased in a unifilar inner coil, the INGEVITY+ leads were upgraded to a trifilar coil design.…”
Section: Methodsmentioning
confidence: 99%
“…These leads incorporate design aspects for MRI conditional safety. The inner coil of the MRI lead was modified to have a higher inductance to prevent heating during MRI scanning, which was achieved through the unifilar design of the inner coils of the INGEVITY leads [ 13 , 15 , 16 ]. However, because torque transfer decreased in a unifilar inner coil, the INGEVITY+ leads were upgraded to a trifilar coil design.…”
Section: Methodsmentioning
confidence: 99%
“…The unique construction of the INGEVITY lead has been previously described 4 . Both the inner and outer conductor coils are a single filar of MP35N wire.…”
Section: Introductionmentioning
confidence: 99%
“…The unique construction of the INGEVITY lead has been previously described. 4 Both the inner and outer conductor coils are a single filar of MP35N wire. This design minimizes the electrical coupling during an MRI in order to reduce potential heating of the tip; however it also reduces the axial strength of the lead when used as a rail for extraction.…”
Section: Introductionmentioning
confidence: 99%
“…he Ingevity family of magnetic resonance imaging provisional pacemaker leads is manufactured by the Boston Scientific Corporation (Marlborough, MA). [1][2][3] Starting in 2017, there were reports of unreliability in the helix deployment mechanism of the Ingevity magnetic resonance imaging lead. 4 This led to lead replacements during implantation procedures.…”
mentioning