2020
DOI: 10.1016/j.apacoust.2020.107374
|View full text |Cite
|
Sign up to set email alerts
|

Reverberation time measurements of a neonatal incubator

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
3
0
1

Year Published

2020
2020
2024
2024

Publication Types

Select...
5
1
1

Relationship

3
4

Authors

Journals

citations
Cited by 8 publications
(7 citation statements)
references
References 42 publications
0
3
0
1
Order By: Relevance
“…Levels were measured twice for each microphone-source combination (i.e., sixteen mic-source measurements). For measurements of the incubator energy distribution, 15 microphone positions were used inside the incubator; this matches the procedure used for small rooms in previous studies conducted by the authors (38,39). A total of 60 noise level measurements were conducted, with four measurements in each of 15 positions.…”
Section: Incubator Acoustic Isolation In Nabs Room (Figure 1b)mentioning
confidence: 99%
“…Levels were measured twice for each microphone-source combination (i.e., sixteen mic-source measurements). For measurements of the incubator energy distribution, 15 microphone positions were used inside the incubator; this matches the procedure used for small rooms in previous studies conducted by the authors (38,39). A total of 60 noise level measurements were conducted, with four measurements in each of 15 positions.…”
Section: Incubator Acoustic Isolation In Nabs Room (Figure 1b)mentioning
confidence: 99%
“…Therefore, in this experiment, the influence of the resonant frequencies of the room can be considered negligible. However, the resonant frequencies of rooms with acoustic reflective surfaces -like NICUs- may influence the T of the incubator, especially considering that the acoustic insulation of the incubator is very poor ( 43 , 44 ).…”
Section: Discussionmentioning
confidence: 99%
“…Inkubator bayi yang baik adalah yang dapat menjaga suhu yang dibutuhkan secara stabil dan merata diseluruh ruang, sehingga inkubator perlu di kalibrasi secara berkala. Ukuran luas inkubator bayi umum sama yaitu panjang ±85cm, lebar ±41 cm, dan tinggi ±40 cm [7] [8]. Pada beberapa kasus ada kemungkinan suhu di dalam inkubator bisa berlebih atau kurang dari normal sehingga pada inkubator bayi umumnya di lengkapi dengan sistem keamanan dan alarm.…”
Section: Iunclassified