2016
DOI: 10.18869/acadpub.ijrr.14.1.73
|View full text |Cite
|
Sign up to set email alerts
|

Response of TLD and RPL personal dosimeters in a national inter-comparison test program

Abstract: Background: Different kinds and models of personal dosimeters are used in individual monitoring by workers. Performance tes ng as part of approval procedures is carried out to demonstrate that the essen al performance specifica ons are rou nely maintained. There are four service providers in Iran which use different luminescence techniques (i. e. TLD and RPL) with various kinds of dosimeter materials/reader instruments in personal dosimetry services. Materials and Methods: A na onal performance approval tests … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
1
0
1

Year Published

2019
2019
2022
2022

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 7 publications
(7 reference statements)
0
1
0
1
Order By: Relevance
“…Atualmente, o princípio da RPL tem sido objeto de estudo para alguns materiais, mas é aplicado comercialmente a apenas um vidro de fosfato, o dosímetro RPL GD-352M (Chiyoda Technol) 21 . Portanto, o maior desafio da expansão da RPL está também na escassez de materiais adequados às diversas aplicações.…”
Section: Dosimetria Luminescente Pessoalunclassified
“…Atualmente, o princípio da RPL tem sido objeto de estudo para alguns materiais, mas é aplicado comercialmente a apenas um vidro de fosfato, o dosímetro RPL GD-352M (Chiyoda Technol) 21 . Portanto, o maior desafio da expansão da RPL está também na escassez de materiais adequados às diversas aplicações.…”
Section: Dosimetria Luminescente Pessoalunclassified
“…Each dosimetry service sends personal dose badges to the National Radiation Standard Laboratory (NRSL) for irradiation following the radiation source category; then then the badges are returned to the original dosimetry services for reading and reporting the personal radiation dose for evaluation. The previous studies related to personal dosimeter measurement accuracy had major discussion on the acceptance comparison between each dosimetry service under current criteria 17 , 18 . However, it is a lack of discussion on forwarding analysis of the factors like years of use and readout frequency, and whether potentially effecting on the measurement accuracy.…”
Section: Introductionmentioning
confidence: 99%