2010
DOI: 10.4085/1062-6050-45.4.338
|View full text |Cite
|
Sign up to set email alerts
|

Response of Thermocouples Interfaced to Electrothermometers When Immersed in 5 Water Bath Temperatures

Abstract: Context: Thermocouples and electrothermometers are used in therapeutic modality research. Until recently, researchers assumed that these instruments were valid and reliable.Objective: To examine 3 different thermocouple types in 56C, 156C, 18.46C, 256C, and 356C water baths.Design: Randomized controlled trial. Setting: Therapeutic modality laboratory. Intervention(s): Eighteen thermocouple leads were inserted through the wall of a foamed polystyrene cooler. The cooler was filled with water. Six thermocouples (… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
11
0

Year Published

2010
2010
2019
2019

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 14 publications
(11 citation statements)
references
References 7 publications
0
11
0
Order By: Relevance
“…The intramuscular thermocouple was cleansed with a mild protein- thermocouples have been studied previously and it has been recommended that researchers test and report on the equipment utilized in the study. 21,22 The three intramuscular thermocouples utilized in the study were pilot tested with the Iso-Thermex electrothermometer and a calibrated mercury thermometer; thermocouples were found to be valid as compared to the mercury thermometer were valid and reliable within 0.1C. Temperatures were recorded every 60-seconds during the 30-minute treatment period.…”
Section: Methodsmentioning
confidence: 99%
“…The intramuscular thermocouple was cleansed with a mild protein- thermocouples have been studied previously and it has been recommended that researchers test and report on the equipment utilized in the study. 21,22 The three intramuscular thermocouples utilized in the study were pilot tested with the Iso-Thermex electrothermometer and a calibrated mercury thermometer; thermocouples were found to be valid as compared to the mercury thermometer were valid and reliable within 0.1C. Temperatures were recorded every 60-seconds during the 30-minute treatment period.…”
Section: Methodsmentioning
confidence: 99%
“…The reliability and validity of the IT-21 thermocouples and Iso-Thermex electrothermometer have been described. 16 We used an imaging ultrasound (LOGIQ 5e; General Electric Company, Fairfield, CT) to measure the subcutaneous fat thickness of our participants and to verify the depth of the inserted thermocouples.…”
Section: Instrumentsmentioning
confidence: 99%
“…Experiment 1 was a 2 3 2 repeated-measures factorial design with extension lead (with, without) and trial (1,2) as the independent variables. Experiment 2 was a 1 3 3 repeated-measures factorial design with phase (preapplication, ice application, postapplication) as the independent variable.…”
Section: Methodsmentioning
confidence: 99%
“…We collected data for each trial every 20 seconds for 5 minutes. 1,2,5 Experiment 2. For Experiment 2, 5 IT-21 and 5 PT-6 thermocouples were set up in our polystyrene cooler similarly to the setup used in experiment 1 but without the use of extension leads.…”
Section: Instrumentsmentioning
confidence: 99%
See 1 more Smart Citation