& A new commercially molecularly imprinted polymer dedicated to the specific solid phase extraction of catecholamines and metanephrines has been tested for the extraction of neurotransmitters in serum samples. A protocol was developed especially for several aminergic neurotransmitters and metabolites (dopamine, 3-methoxytyramine, epinephrine, norepinephrine, normetanephrine, and serotonine).The serum was diluted in water (1=10, v=v), loaded on a cartridge filled with molecularly imprinted polymer and washed successively with water, methanol-water (15=85, v=v), and methanol. Analytes were then eluted with methanol-acetic acid (99=1, v=v). No matrix peak was observed on the chromatograms obtained by high performance liquid chromatography coupled with ultra-violet detection (281 nm). High selectivities were observed between imprinted and nonimprinted polymers. Extraction recoveries ranged from 64-98% (RSD <6%, n ¼ 4) depending on the type of analytes and linearity was assessed in the 0.30-0.90 lM concentration range by least squares regression (r 2 > 0.991).Then, large volumes of sample (8-40 mL) were loaded on the cartridge to evaluate the binding capacity of the cartridge. Increasing loaded sample volumes induced a decrease in extraction recovery, particularly for hydrophilic compounds. For a given elution recovery, higher preconcentration factors were achieved ($55) for less hydrophilic compounds (sample volume of 40 mL).