“…Two-dimensional SAXS intensity was measured by P45 phosphor coupled with tapered optics (FOT) and a CCD detector cooled down to 233 K. Detail of the measurement system is given in 26) , where the measurements have been made at the K absorption edge of Al for the samples that have strong scattering contrast with isotropic profile. Calibration of scattering angle was made by the Debye rings of silver behenate 26,27) , where isotropic rings with uniform intensity along the rings were observed, showing that distortion correction is not necessary. Although several metastable phases are reported for Al-Mg binary alloys 28,29) , only one isolated dissolution peak was observed, which was attributed to GP zones.…”