Radiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film

Abstract: Micro‐multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation system and requires the use of detectors with high spatial resolution, high sensitivity and practically no energy depende… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

2
38
0

Year Published

2011
2011
2020
2020

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 45 publications
(40 citation statements)
references
References 15 publications
2
38
0
Order By: Relevance
“…The Novalis' mMLC has a consistently smaller penumbra for the circular fields and compared to the Synergy-S' MLC. It is to be noted that the measured penumbra for Novalis (BrainLAB) closely matched with that reported by Clark et al [4], Chern et al [5], and García-Garduño et al [8]. The effective penumbras for these circular fields range from about 4 AE 0.15 mm to 7 AE 0.27 mm for the Synergy-S mMLC (Fig.…”
Section: Results and Discussion Penumbrasupporting
confidence: 72%
See 1 more Smart Citation
“…The Novalis' mMLC has a consistently smaller penumbra for the circular fields and compared to the Synergy-S' MLC. It is to be noted that the measured penumbra for Novalis (BrainLAB) closely matched with that reported by Clark et al [4], Chern et al [5], and García-Garduño et al [8]. The effective penumbras for these circular fields range from about 4 AE 0.15 mm to 7 AE 0.27 mm for the Synergy-S mMLC (Fig.…”
Section: Results and Discussion Penumbrasupporting
confidence: 72%
“…Chern et al [5] reported that the 3 mm mMLC of BrainLAB provided better target conformity and greater normal tissue sparing than the 5 mm MLC in stereotactic radiosurgery using dynamic conformal arcs. García-Garduño et al [8] have recently reported penumbra (2.26 AE 0.11 mm along leaf side and 2.31 AE 0.11 mm along lead end), transmission (0.93 AE 0.05%), and leakage (1.08 AE 0.08%) of m 3 -mMLC of BrainLAB for 6 MV photon beam study using EBT film.…”
Section: Introductionmentioning
confidence: 97%
“…Asnaashari et al 16 assessed two MLCs for radiosurgery and reported penumbrae in the range of 3.5-4.8 mm. Fischer et al 17 and García-Garduño et al 18 determined penumbrae of around 2.5 mm, in a slightly different measurement setup. Sharma et al 19 found a range of 2-4 mm for 6 MV delivery modes and different field sizes, and Solberg et al 20 reported around 4.5 mm for a 10×10 cm 2 field on a gimbaled Linac.…”
Section: Discussionmentioning
confidence: 98%
“…Since its introduction in 2004, Gafchromic's EBT radiochromic film (International Specialty Products, Wayne, USA) has been widely adopted for use in quality assurance testing of clinical radiotherapy treatments including 3D conformal [1], intensity modulated and stereotactic treatments [2][3][4][5] as well as arc therapies [6] including tomotherapy [7]. EBT has been rigorously investigated and found to be capable of producing reliable and reproducible dose measurements, with a high degree of spatial accuracy and uniformity of response [1,[8][9][10][11][12][13].…”
Section: Introductionmentioning
confidence: 99%