Radiation Thermometry and Emissivity Measurements Under Vacuum at the PTB

Abstract: A new experimental facility was realized at the PTB for reducedbackground radiation thermometry under vacuum. This facility serves three purposes: (i) providing traceable calibration of space-based infrared remote-sensing experiments in terms of radiation temperature from −173 • C to 430 • C and spectral radiance; (ii) meeting the demand of industry to perform radiation thermometric measurements under vacuum conditions; and (iii) performing spectral emissivity measurements in the range from 0 • C to 430 • C wi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
45
0

Year Published

2008
2008
2020
2020

Publication Types

Select...
3
3
1

Relationship

2
5

Authors

Journals

citations
Cited by 78 publications
(45 citation statements)
references
References 8 publications
0
45
0
Order By: Relevance
“…At the Physikalisch-Technische Bundesanstalt (PTB), a reduced-background calibration facility (RBCF) [1][2][3][4] was constructed. The RBCF is intended for the calibration of cryogenic and near-ambient-temperature blackbodies and infrared detector systems under medium vacuum ( p < 10 −2 Pa) or controlled atmosphere conditions and reduced-background radiation (liquid-nitrogencooled radiation shields).…”
Section: Introductionmentioning
confidence: 99%
“…At the Physikalisch-Technische Bundesanstalt (PTB), a reduced-background calibration facility (RBCF) [1][2][3][4] was constructed. The RBCF is intended for the calibration of cryogenic and near-ambient-temperature blackbodies and infrared detector systems under medium vacuum ( p < 10 −2 Pa) or controlled atmosphere conditions and reduced-background radiation (liquid-nitrogencooled radiation shields).…”
Section: Introductionmentioning
confidence: 99%
“…The Emissivity Measurement in Air Facility (EMAF) allows measurements in the temperature range from 25 °C to 500 °C and in a wavelength range from 1.5 µm to 50 µm under angles of observation ranging from 0° to 80°. The Reduced Background Calibration Facility (RBCF) [3] is used for radiation thermometry measurements under vacuum in the temperature range from -170 °C to 430 °C and emissivity measurements from -40 °C to 1000 °C. The RBCF (see Figure 5) is the ideal instrument for the calibration of remote sensing applications avoiding any atmospheric interferences [4].…”
Section: Emissivity Measurementsmentioning
confidence: 99%
“…The spectral distribution of uncertainty has to be calculated for each individual measurement via a Monte-Carlo based method which is described in detail in [3].…”
Section: Emissivity Measurementsmentioning
confidence: 99%
See 1 more Smart Citation
“…Using a newly developed facility for radiation thermometry under vacuum at PTB [2][3][4] directional spectral emissivity, total directional emissivity and total hemispherical emissivity measurements in a broad wavelength and temperature range can be performed highly accurate under vacuum with traceable uncertainties.…”
Section: Introductionmentioning
confidence: 99%