Quantitative methanol-burning lung model for validating gas-exchange measurements over wide ranges of F I O 2

Abstract: The methanol-burning lung model has been used as a technique for generating a predictable ratio of carbon dioxide production (VCO2) to oxygen consumption (VO2) or respiratory quotient (RQ). Although an accurate RQ can be generated, quantitatively predictable and adjustable VO2 and VCO2 cannot be generated. We describe a new burner device in which the combustion rate of methanol is always equal to the infusion rate of fuel over an extended range of O2 concentrations. This permits the assembly of a methanol-burn… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
15
0

Year Published

2005
2005
2022
2022

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 15 publications
(16 citation statements)
references
References 18 publications
(15 reference statements)
1
15
0
Order By: Relevance
“…Percent recoveries of both O 2 and CO 2 , as well as RER, were calculated by previously published formulas . The theoretical value for percent O 2 and CO 2 recovery was 100%, and the theoretical value for RER was 0.667 based on the ratio of produced CO 2 to utilized O 2 in the burning of methanol from this equation: 2CH 3 OH + 3O 2 → 2CO 2 + 4H 2 O, which is the value regardless of fuel infusion rate and/or changes in fraction of inspired oxygen (F io 2 ) …”
Section: Methodsmentioning
confidence: 99%
“…Percent recoveries of both O 2 and CO 2 , as well as RER, were calculated by previously published formulas . The theoretical value for percent O 2 and CO 2 recovery was 100%, and the theoretical value for RER was 0.667 based on the ratio of produced CO 2 to utilized O 2 in the burning of methanol from this equation: 2CH 3 OH + 3O 2 → 2CO 2 + 4H 2 O, which is the value regardless of fuel infusion rate and/or changes in fraction of inspired oxygen (F io 2 ) …”
Section: Methodsmentioning
confidence: 99%
“…Midownik et al (2000) developed a semi‐automated gas exchange measurement system based on a Douglas bag design. This system was validated using a modified version of the quantitative funnel burner lung model (Miodownik et al, 1998) and standard water volume displacement. They compared their validated system to the Sensor Medics Deltatrac (mixing chamber Yorba Linda, California, USA) and Medical Graphics CPX (breath‐by‐breath) systems (Medical Graphics Corp, St. Paul, MN, USA).…”
Section: Comparison Of One or More Automated Metabolic Analysis Systementioning
confidence: 99%
“…In the study of McClave et al [26] steady state criterion is defined as a five consecutive 1-min intervals where VO 2 and VCO 2 change by\10% to produce the best correlation and smallest mean difference within a 24-h EE evaluation. Though the criterion of McClave et al [26] is quiet stringent, patients who do not achieve the steady state criterion are considered as clinically unstable and the measurement value cannot be extrapolated for the 24-h TEE [29]. In the present study of 77 subjects, only 42 subjects achieved steady state criterion, denoting a high participant loss.…”
Section: Comparison Of the Indirect Calorimetersmentioning
confidence: 56%
“…In extreme obese patients, especially those who have a BMI higher than 40, EE is highly variable [29], and the only known accurate way in optimising nutritional support regimes is through ICs [15]. Therefore, the assimilation of ICs is vital to define, if between instruments RMR can be accurately estimated.…”
Section: Comparison Of the Indirect Calorimetersmentioning
confidence: 99%