Quantification of chloride ion impurities in ionic liquids using ICP-MS analysis

Abstract: The residual chloride ion content in 1-butyl-3-methylimidazolium tetrafluoroborate, [bmim]BF 4 , prepared via various methods, was analysed by ICP-MS. A preliminary analysis was also performed on 1-butyl-3-methylimidazolium hexafluorophosphate, [bmim]PF 6 .

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
21
0
1

Year Published

2005
2005
2017
2017

Publication Types

Select...
6
3

Relationship

0
9

Authors

Journals

citations
Cited by 33 publications
(25 citation statements)
references
References 18 publications
0
21
0
1
Order By: Relevance
“…Conductivity measurements were carried out using a Metrohm 660 conductivity meter equipped with two Pt black electrodes, with a cell constant of 0.852 cm -1 . Chloride analysis of the RTILs was determined according to the method used by McCamley et al [20] with a PerkinElmer Sciex Elan 6100 ICP-MS. Finally, water in the RTILs was determined using a Karl Fisher model 701 KF Titrino system with Hydranal Composite 5 as titrant and HPLC grade methanol as the base.…”
Section: Further Analytical Techniquesmentioning
confidence: 99%
“…Conductivity measurements were carried out using a Metrohm 660 conductivity meter equipped with two Pt black electrodes, with a cell constant of 0.852 cm -1 . Chloride analysis of the RTILs was determined according to the method used by McCamley et al [20] with a PerkinElmer Sciex Elan 6100 ICP-MS. Finally, water in the RTILs was determined using a Karl Fisher model 701 KF Titrino system with Hydranal Composite 5 as titrant and HPLC grade methanol as the base.…”
Section: Further Analytical Techniquesmentioning
confidence: 99%
“…Ainsi McCamley et al (2004) ont utilisé un ICP-MS haute résolution (ICP-HRMS) pour mesurer les impuretés de chlore résiduel après synthèse dans 6 liquides ioniques issus de voies de synthèse différentes. Avec une limite de quantification de 10 µg/l, le gain en sensibilité de l'ICP-HRMS au regard des techniques de type électrophorèse capillaire est de 100 fois.…”
Section: Detection Limit With Different Injection Device In Icp-msunclassified
“…11,12 Moreover, results can be questionable in case of hydrophobic ionic liquids, as these methods typically rely on aqueous media. 13 Optimized detection methods such as chloride sensors, 6 ion chromatography [14][15][16] X-ray uorescence, 17 capillary electrophoresis, 18 cathodic stripping voltammetry, 19 and even inductively coupled plasma mass spectrometry (ICP-MS) 20 have been reported to be employed for residual chloride determination. These methods offer higher sensitivity and accuracy compared to classical wet chemical methods; however, they lack in ease of use and oen require sophisticated sample preparation.…”
Section: Introductionmentioning
confidence: 99%