2008
DOI: 10.1111/j.1442-2042.2007.01934.x
|View full text |Cite
|
Sign up to set email alerts
|

Prevention of urethral stricture in insertion of an inflatable penile prosthesis

Abstract: Objective: The aim of this study was to find the mechanism that causes urethral stricture with a view to preventing its occurrence after implantation of a three-piece inflatable penile prosthesis. Methods: When implanting a three-piece inflatable penile prosthesis, we draw two longitudinal lines parallel to the long axis of the proximal cylinder and the tubing leading from each cylinder into the cavernosal space, to prevent 360 degree rotation of the cylinder. Results: From September 1993 to February 2007, a t… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
8
0

Year Published

2008
2008
2022
2022

Publication Types

Select...
5
1

Relationship

2
4

Authors

Journals

citations
Cited by 7 publications
(8 citation statements)
references
References 24 publications
0
8
0
Order By: Relevance
“…Authors described a high complication rate of urethral stricture of 5.8%, although infection has not occurred, hence, authors should add an explanation for the current high complication rate. The events reported in this article 3 are extremely rare and have not been reported by any other institutions.…”
mentioning
confidence: 64%
“…Authors described a high complication rate of urethral stricture of 5.8%, although infection has not occurred, hence, authors should add an explanation for the current high complication rate. The events reported in this article 3 are extremely rare and have not been reported by any other institutions.…”
mentioning
confidence: 64%
“…To overcome the abnormal rotation, we marked two longitudinal lines on the surface of the cylinder and tubings leading from the cylinders before the insertion of cylinder of TPIPP [9].…”
Section: Discussionmentioning
confidence: 99%
“…After urethroplasty, new cylinders and pump were placed in cavernosal spaces and intra-scrotum in a routine manner, respectively ( Figure 2B). We have drawn two longitudinal lines on the cylinder and tubings leading from cylinder to prevent the abnormal rotation of the cylinder [9]. Intraoperative antibiotic irrigation (gentamycin and vancomycin) was performed in all procedures.…”
Section: Case Studymentioning
confidence: 99%
See 1 more Smart Citation
“…Urethral stricture after implantation of a three-piece inflatable penile prosthesis is very rare but has been reported in 2007 [10]. Urethral stricture was induced by compression of twisted tubing in three-piece inflatable penile prosthesis and it would be aggravated with the lapse of time [11].…”
Section: Fig 1 Coloplast Titan Inflatable Penile Prosthesismentioning
confidence: 99%