Pressurized bag pump and syringe pump arterial flushing systems: an unrecognized hazard in neonates?

Abstract: Based on data about critical flow velocities through an radial arterial cannula in neonates, both tested flushing systems carry the risk of exceeding the critical value of 0.5 ml/s. They are likely to cause retrograde embolization of flushing solution into the central arterial circulation with the associated risk of clot and air embolization. In vivo studies should identify margins of safety to minimize the risk of retrograde flushing into the central arterial circulation.