2003
DOI: 10.1007/bf03021037
|View full text |Cite
|
Sign up to set email alerts
|

Arterial fast bolus flush systems used routinely in neonates and infants cause retrograde embolization of flush solution into the central arterial and cerebral circulation

Abstract: P Pu ur rp po os se e: : To evaluate the risk of retrograde embolization of flush solution in neonates and infants with routinely used electronic syringe pumps and infusion bag pump flush systems.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
4
0
2

Year Published

2003
2003
2023
2023

Publication Types

Select...
3
3

Relationship

0
6

Authors

Journals

citations
Cited by 6 publications
(6 citation statements)
references
References 26 publications
0
4
0
2
Order By: Relevance
“…A volume of air of 1.6 mL injected in the radial artery under high pressure was sufficient to evoke neurologic and hemodynamic reactions in an adult patient. It has been demonstrated in infants that rapid flushing not only results in retrograde embolization of flush solution, but also causes local arterial vasospasm, transient elevation of arterial blood pressure and intracranial/intraventricular pressure [ 22 ]. This can be related to the small size of the patient, but also to the immaturity of cardiovascular and nervous system.…”
Section: Discussionmentioning
confidence: 99%
“…A volume of air of 1.6 mL injected in the radial artery under high pressure was sufficient to evoke neurologic and hemodynamic reactions in an adult patient. It has been demonstrated in infants that rapid flushing not only results in retrograde embolization of flush solution, but also causes local arterial vasospasm, transient elevation of arterial blood pressure and intracranial/intraventricular pressure [ 22 ]. This can be related to the small size of the patient, but also to the immaturity of cardiovascular and nervous system.…”
Section: Discussionmentioning
confidence: 99%
“…However, the two accompanying papers in the Journal 10,11 confirm that as our specialty changes adverse outcomes still occur, often in new forms.…”
mentioning
confidence: 99%
“…10 This is an example of how monitoring used to improve patient care is accompanied by the risk of adverse events.…”
mentioning
confidence: 99%
See 2 more Smart Citations