Polarized Photometric Detector for High-Performance Liquid Chromatography

Abstract: A photometric detector for an optically active compound is described in which two polarizers are set on both sides of the flow cell. When a nonchromophoric, optically active compound passes through the cell, the intensity of natural light transmitted through the polarizers and flow cell is changed. The detection of such a sample is possible photometrically.A large inclination angle between the polarizers increases the absorbance change and allows dextrorotatory and levorotatory to be differentiated.The direct … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
6
0

Year Published

1994
1994
2011
2011

Publication Types

Select...
6

Relationship

3
3

Authors

Journals

citations
Cited by 16 publications
(6 citation statements)
references
References 5 publications
0
6
0
Order By: Relevance
“…In addition, any absorption of the sample at the selected wavelength will induce a reading offset, which is often much larger than the effect to be measured, and consequently difficult to compensate. 7 Alternative approaches based on the use of two polarizers and changing their relative orientation in the sample compartment of a single beam spectrophotometers were reported many years ago. These limiting factors may be due to some instability of the light source, which is not compensated in a single beam experiment.…”
Section: Intensity Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…In addition, any absorption of the sample at the selected wavelength will induce a reading offset, which is often much larger than the effect to be measured, and consequently difficult to compensate. 7 Alternative approaches based on the use of two polarizers and changing their relative orientation in the sample compartment of a single beam spectrophotometers were reported many years ago. These limiting factors may be due to some instability of the light source, which is not compensated in a single beam experiment.…”
Section: Intensity Methodsmentioning
confidence: 99%
“…A similar setup has been reported recently 6 with a diode array spectrophotometer, while the same approach had been devised in the past for high performance liquid chromatography (HPLC) OR detectors. 7 Alternative approaches based on the use of two polarizers and changing their relative orientation in the sample compartment of a single beam spectrophotometers were reported many years ago. 8,9 In the apparatus by Keston et al, which Very similar values are obtained by use of eq.…”
Section: Intensity Methodsmentioning
confidence: 99%
“…We first developed a non-modulated OR detector. 1,2 We then developed a more sensitive CD detector 3 and a fluorescence-detected CD (FDCD) spectrometer, 4 which used a modulated beam in the wavelength axial direction. As is obvious from comparisons between OR, CD, and FDCD, the sensitivities are in the order OR < CD < FDCD whereas the number of compounds that can be analyzed is in the reverse order.…”
Section: Introductionmentioning
confidence: 99%
“…We have proposed a novel, non-modulated polarimetric detector, a polarized photometric detector (PPD) (Yamamoto et al, 1991), and have continued research on its fundamental theory and applications Yamamoto et al, 1996;Yamamoto et al, 1997). In these processes, we have developed a split-type flow cell for the purpose of removing the influences of absorbance and/or RI changes followed by the analyte elution (Yamamoto et al, 1995).…”
Section: Introductionmentioning
confidence: 99%