Pacemaker Tachycardia: Is it Pacemaker‐Mediated Tachycardia or Sinus Tachycardia?

Abstract: A 38-year-old woman received a transvenous St. Jude Atlas + HF V-340 biventricular implantable cardioverter defibrillator (ICD; Sylmar, CA, USA) for a nonischemic dilated cardiomyopathy associated with drug-refractory congestive heart failure, left bundle branch block, and a left ventricular ejection fraction of 20%. About a year later she was referred to our center for the first time after sustaining 2 shocks related to supraventricular tachycardia at a rate of 210 bpm, which was documented in stored electrog… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
2
0

Year Published

2010
2010
2012
2012

Publication Types

Select...
3
1

Relationship

2
2

Authors

Journals

citations
Cited by 4 publications
(2 citation statements)
references
References 2 publications
0
2
0
Order By: Relevance
“…The mathematical constraints of the traditional PMT detection algorithm make a false positive response of a PMT highly unlikely. However, rare cases of sinus tachycardia have been reported to activate the PMT termination algorithm . In the present case, the episodes of Wenckebach upper rate behavior ruled out the clinical diagnosis of PMT.…”
Section: Discussionmentioning
confidence: 57%
“…The mathematical constraints of the traditional PMT detection algorithm make a false positive response of a PMT highly unlikely. However, rare cases of sinus tachycardia have been reported to activate the PMT termination algorithm . In the present case, the episodes of Wenckebach upper rate behavior ruled out the clinical diagnosis of PMT.…”
Section: Discussionmentioning
confidence: 57%
“…However, rare cases of sinus tachycardia have been reported to activate the PMT termination algorithm. 2,3 Figures 2A and B show a PMT in a patient with a St. Jude Promote implantable cardioverterdefibrillator (ICD). 1 At the time of follow-up, stored data revealed a PMT that was terminated automatically by the PMT-terminating algorithm but the test AV delay lengthened by about 50 ms.…”
mentioning
confidence: 99%