“…The following multiple reaction monitoring (MRM) pairs were utilized for the analysis of each SQOOH isomer: m/z 465.4 > 433.5, 2‐OOH‐SQ; m/z 465.4 > 395.3, 3‐OOH‐SQ; m/z 465.4 > 365.3, 6‐OOH‐SQ; m/z 465.4 > 297.2, 10‐OOH‐SQ; m/z 465.4 > 181.1, 7‐OOH‐SQ; m/z 465.4 > 229.2, 11‐OOH‐SQ. Other MS parameters, consistent with our previous study, were as follows: curtain gas, 30.0 psi; collision gas, 7.0 psi; ion spray voltage, 5500.0 V; temperature, 400.0 °C; ion source gas 1, 80.0 psi; and ion source gas 2, 60.0 psi. External calibration curves for quantitation were prepared with the use of synthetic SQOOH isomer standards.…”