2016
DOI: 10.1109/tbme.2016.2570125
|View full text |Cite
|
Sign up to set email alerts
|

On the merits of tetrapolar impedance spectroscopy for monitoring lithium concentration variations in human blood plasma

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
18
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
6
1
1

Relationship

1
7

Authors

Journals

citations
Cited by 11 publications
(18 citation statements)
references
References 36 publications
(35 reference statements)
0
18
0
Order By: Relevance
“…Samples of human plasma that have been spiked with lithium were prepared by mixing 1 mL from each reagent of Li2CO3 prepared earlier i.e. 18,24,30,36,42,48,54 and 60 mEq with 29 mL of blood plasma, thus giving 8 samples of 30 mL, with lithium concentrations ranging between 0.4-2.0 mEq. To check for repeatability of measurement, another set of samples were prepared in the same way using a different batch of blood plasma, as well as an additional set of samples spiked with unknown lithium concentrations in plasma.…”
Section: A Preparation Of Blood Samplesmentioning
confidence: 99%
See 2 more Smart Citations
“…Samples of human plasma that have been spiked with lithium were prepared by mixing 1 mL from each reagent of Li2CO3 prepared earlier i.e. 18,24,30,36,42,48,54 and 60 mEq with 29 mL of blood plasma, thus giving 8 samples of 30 mL, with lithium concentrations ranging between 0.4-2.0 mEq. To check for repeatability of measurement, another set of samples were prepared in the same way using a different batch of blood plasma, as well as an additional set of samples spiked with unknown lithium concentrations in plasma.…”
Section: A Preparation Of Blood Samplesmentioning
confidence: 99%
“…Therefore, variations in the recorded voltage can directly translate to blood lithium concentration changes. A more detailed explanation of the method can be found in [24], where the sensitivity of the method to changes in lithium concentration in human blood The new probe features a tenfold increase in the total sensitivity, corresponding to a tenfold increase in the measured impedance change for specific Li concentration variations. Hence, with appropriately designed instrumentation for tetrapolar impedance measurements, the new probe has the potential to measure very small Li concentration variations.…”
Section: Tetrapolar Electrical Impedance Measurements Of Blood Litmentioning
confidence: 99%
See 1 more Smart Citation
“…Lithium carbonate was selected as the analyte since it is a very interesting and challenging PoC application. Multiple approaches have been used to try and detect it in blood due to its use as a therapeutic means for multiple psychiatric disorders [21][22][23]. The lithium carbonate solutions are made using distilled water and have a very low concentration.…”
Section: Phononic Crystal Sensormentioning
confidence: 99%
“…Saline was used as a test reagent because its concentration has a very linear relationship with the measured impedance. (6,28) In general, it is cost effective to achieve accuracy using calibration instead of making sensors infinitely precise. Calibration types can be categorized as individual calibration, in which parameters are applied to individual sensors, and lot calibration, in which the same parameters are applied to the entire production lot.…”
Section: Introductionmentioning