2011
DOI: 10.1117/12.894659
|View full text |Cite
|
Sign up to set email alerts
|

NuSTAR ground calibration: The Rainwater Memorial Calibration Facility (RaMCaF)

Abstract: The Nuclear Spectroscopic Telescope Array (NuSTAR) is a NASA Small Explorer mission that will carry the first focusing hard X-ray (5 − 80 keV ) telescope to orbit. The ground calibration of the three flight optics was carried out at the Rainwater Memorial Calibration Facility (RaMCaF) built for this purpose. In this article we present the facility and its use for the ground calibration of the three optics.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
13
0

Year Published

2011
2011
2020
2020

Publication Types

Select...
6
2

Relationship

3
5

Authors

Journals

citations
Cited by 16 publications
(13 citation statements)
references
References 20 publications
0
13
0
Order By: Relevance
“…An overview of its use for the NuSTAR flight optics calibration is presented in these proceedings (Brejnholt et al, "NuSTAR ground calibration: The Rainwater Memorial Calibration Facility (RaMCaF)", SPIE 8147- 16, 2011). 17 It describes the facility setup for the NuSTAR flight optics and an overview of the types of calibration measurements that were performed. It also illustrates the optics alignment procedure and modeling of small mounting distortions that occurred during the ground calibration which are important to second order when evaluating the ground calibration data, but will not be present in space.…”
Section: X-ray Calibration Measurements At the Rainwater Memorial Calmentioning
confidence: 99%
See 1 more Smart Citation
“…An overview of its use for the NuSTAR flight optics calibration is presented in these proceedings (Brejnholt et al, "NuSTAR ground calibration: The Rainwater Memorial Calibration Facility (RaMCaF)", SPIE 8147- 16, 2011). 17 It describes the facility setup for the NuSTAR flight optics and an overview of the types of calibration measurements that were performed. It also illustrates the optics alignment procedure and modeling of small mounting distortions that occurred during the ground calibration which are important to second order when evaluating the ground calibration data, but will not be present in space.…”
Section: X-ray Calibration Measurements At the Rainwater Memorial Calmentioning
confidence: 99%
“…Further, a slight temporary asymmetry in the PSF is present for the ground calibration measurements due to gravitational sag (~1 mdeg) and the alignment struts introducing a parabolic bend in the optic mandrel (~2 mdeg) that acts over all of the mirror layers. 17 The fit range of -25 < X < 15 mm was used to avoid contamination of single bounce Xrays. The HPD was calculated from the fit parameters for an infinite focal plane.…”
Section: One-dimensional Profile Measurements Of the Psf For Each 60 mentioning
confidence: 99%
“…Facility performance [17] and initial results from the flight calibrations [18] will be presented in August 2011 at the SPIE Optics + Photonics in San Diego, CA.…”
Section: Ramcaf Now and Tomorrowmentioning
confidence: 99%
“…The strength of the on-ground calibration data set, compared to in-orbit data, derives from these subgroup measurements, where only a small fraction of the optic is investigated, effectively improving the resolution of the ORM validation. Detailed accounts of the calibration approach and hardware are available from Brejnholt et al 2 , Brejnholt et al 3 and Koglin et al 9 .…”
Section: On-ground Calibrationmentioning
confidence: 99%
“…The on-ground calibration of the two multilayer coated flight optics was carried out at the Rainwater Memorial Calibration Facility for X-ray optics (RaMCaF) 2,3 at Nevis Laboratories in New York. The multilayer coating was laid down at the Technical University of Denmark (DTU Space) 4 .…”
Section: Introductionmentioning
confidence: 99%