2012
DOI: 10.1063/1.3681442
|View full text |Cite
|
Sign up to set email alerts
|

Note: Absolute calibration of two DRZ phosphor screens using ultrashort electron bunch

Abstract: This article gives the absolute calibration of two types phosphor screens (DRZ) that were used to detect and characterize electron bunches driven by laser-plasma accelerator. The test was performed with picoseconds electron bunch at a radio frequency linear electron accelerator in Tsinghua University. The photons emitted from DRZ screens showed good linear responses to the charge of incident electron bunch and cosine angular distribution in space. An energy conversional efficiency of effective scintillant matt… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

1
21
0

Year Published

2012
2012
2024
2024

Publication Types

Select...
9

Relationship

0
9

Authors

Journals

citations
Cited by 29 publications
(22 citation statements)
references
References 13 publications
1
21
0
Order By: Relevance
“…Simulations show that the energy deposition of the electrons inside the photo-luminescent layer is almost independent of their kinetic energy above a threshold value of 3 MeV. 16,19,22,23 Additionally, Nakamura et al 21 demonstrated that the sensitivity of Lanex Fast to highly relativistic electrons has only a weak dependency of the electron energy. It decreases by 1% per 100 MeV increase of the electron energy.…”
Section: Methodsmentioning
confidence: 99%
“…Simulations show that the energy deposition of the electrons inside the photo-luminescent layer is almost independent of their kinetic energy above a threshold value of 3 MeV. 16,19,22,23 Additionally, Nakamura et al 21 demonstrated that the sensitivity of Lanex Fast to highly relativistic electrons has only a weak dependency of the electron energy. It decreases by 1% per 100 MeV increase of the electron energy.…”
Section: Methodsmentioning
confidence: 99%
“…In order to estimate the energy of the dispersed electron beams, knowledge of the undeflected path is required, which was obtained from the correlation between the electron beam pointing (recorded without a magnet) on two phosphor screens (phosphor 1 and phosphor 2). The electron beam charge was estimated using standard calibration data available for the various phosphor screens used [33][34][35].…”
Section: Experimental Setup and Generationmentioning
confidence: 99%
“…The low-energy cutoff of the electron spectrometer was 10 MeV. The electron beam charge was estimated using the absolute calibration data of the phosphor screen [13]. The electron beam profile and the pointing variation were measured by moving the magnet out from the path of the electron beam.…”
mentioning