Abstract: To address the quality assurance (QA) of a MR-linac which is an MRI combined with a linear accelerator (linac), the traditional linac QA-tests need to be redesigned, since the presence of the static magnetic field in the MR-linac alters the electron trajectory. The latter causes the asymmetry in the dose kernel which is introduced by the magnetic field and hinders accurate geometrical QA-tests for the MR-linac. We introduced the use of electron dense materials (e.g. copper) to reduce the size of the dose kerne… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
15
0

Year Published

2016
2016
2024
2024

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 23 publications
(15 citation statements)
references
References 12 publications
0
15
0
Order By: Relevance
“…This model includes the treatment beam with a nominal beam energy of 7 MV, the MLC as well as the cryostat. Each of the 160 leaves of the MLC, which is fixed to 90° (coordinate system according to IEC-61217), has an isocenter-projected width of 7.15 mm and the maximum field size is 27.20 cm parallel to leaf direction and 57.15 cm perpendicular to leaf direction [21] . This large field of view allows treatment of peripheral tumors even though the treatment beam isocenter is fixed to the center of the bore.…”
Section: Methodsmentioning
confidence: 99%
“…This model includes the treatment beam with a nominal beam energy of 7 MV, the MLC as well as the cryostat. Each of the 160 leaves of the MLC, which is fixed to 90° (coordinate system according to IEC-61217), has an isocenter-projected width of 7.15 mm and the maximum field size is 27.20 cm parallel to leaf direction and 57.15 cm perpendicular to leaf direction [21] . This large field of view allows treatment of peripheral tumors even though the treatment beam isocenter is fixed to the center of the bore.…”
Section: Methodsmentioning
confidence: 99%
“…This will eliminate any systematic offsets relating from moveable collimator components. For a spoke film, the use of copper sheets 30,73 removes the dose shift caused by the magnetic field (see section 2.3). This results in the spokes effectively representing the photon beam trajectory.…”
Section: C4 Radiation Isocentermentioning
confidence: 99%
“…This technique has been used to measure profiles and star shot films allowing for common QA techniques to be used from conventional radiotherapy systems with only minor modifications to the methods. 73…”
Section: C Function and Dose Response Of Detectorsmentioning
confidence: 99%
“…3c). Due to anatomical changes stemming from patient movement (such as breathing or digestion), imaging is ideally conducted in real time 229 ; in fact, the use of advanced real-time IGRT is rapidly growing through the development of MR-LINAC [230][231][232][233] and PET-LINAC 234 systems.…”
Section: Imaging and Radiotherapymentioning
confidence: 99%