Measurement, simulation and uncertainty assessment of implant heating during MRI

Abstract: The heating of tissues around implants during MRI can pose severe health risks, and careful evaluation is required for leads to be labeled as MR conditionally safe. A recent interlaboratory comparison study has shown that different groups can produce widely varying results (sometimes with more than a factor of 5 difference) when performing measurements according to current guidelines. To determine the related difficulties and to derive optimized procedures, two different generic lead structures have been inves… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
73
0

Year Published

2010
2010
2019
2019

Publication Types

Select...
5
2
1

Relationship

0
8

Authors

Journals

citations
Cited by 81 publications
(73 citation statements)
references
References 17 publications
(20 reference statements)
0
73
0
Order By: Relevance
“…2) Measurements setup-Data of ||E ⃗ || and ||B ⃗ || were collected using a robotic measurement system (DASY 5NEO, SPEAG, Zurich, Switzerland) [22,38,39] with two ||E ⃗ || probes (ER3DV6 and EX3DV4 for measurements in air and saline, respectively) and a ||H ⃗ || probe (H3DV7) (SPEAG, Zurich, Switzerland). For each measurement point the probes returned three RMS values -one for each field component x, y, and z.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…2) Measurements setup-Data of ||E ⃗ || and ||B ⃗ || were collected using a robotic measurement system (DASY 5NEO, SPEAG, Zurich, Switzerland) [22,38,39] with two ||E ⃗ || probes (ER3DV6 and EX3DV4 for measurements in air and saline, respectively) and a ||H ⃗ || probe (H3DV7) (SPEAG, Zurich, Switzerland). For each measurement point the probes returned three RMS values -one for each field component x, y, and z.…”
Section: Methodsmentioning
confidence: 99%
“…The distributed capacitors present in the physical coil were modeled as two conductive rectangular slabs connected by numerically defined lumped elements, as in [22]. The two baluns present at the physical sources were not modeled in the numerical coils.…”
Section: B Computational Modelingmentioning
confidence: 99%
See 1 more Smart Citation
“…The system is installed at the Department of Technology and Health of the Italian National Institute of Health and has been specially designed to provide a controlled exposure for MRI safety testing with regards to the RF field. Systems with the same features have been widely used in previous published studies (17,(22)(23)(24). It has also been demonstrated that, given the same implant configuration and the same WB-SAR delivered to the phantom, the induced heating inside the isolated birdcage coil is comparable with the one obtained in an actual MRI system using clinical MRI sequences (23).…”
Section: Exposure Systemmentioning
confidence: 88%
“…11d) (Bassen et al 2006;Neufeld et al, 2009). In particular Neufeld et al, (2009) realized their catheter staring from a 0.6 mm thick wire wound to form a helix and including a 10-mm long straight section. The helical wire is insulated by a thin coating (0.012 mm).…”
Section: Catheter Modelsmentioning
confidence: 99%