Manual ventilation devices in neonatal resuscitation: Tidal volume and positive pressure-provision

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

9
45
1
10

Year Published

2011
2011
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 73 publications
(65 citation statements)
references
References 20 publications
9
45
1
10
Order By: Relevance
“…17,18 The present study shows that the high variability does not seem to be an artifact due to restricted sample size, because it was still observed even with a considerably greater number of volunteers (n ϭ 172). This is in agreement with the results of Roehr et al, 25 in which high variability of PIP (2.7-58.9 cm H 2 O) and V T (1-14.6 mL) was observed in a sample of 120 volunteers during ventilation of a neonatal mannequin with a self-inflating bag.…”
Section: Discussionsupporting
confidence: 93%
See 2 more Smart Citations
“…17,18 The present study shows that the high variability does not seem to be an artifact due to restricted sample size, because it was still observed even with a considerably greater number of volunteers (n ϭ 172). This is in agreement with the results of Roehr et al, 25 in which high variability of PIP (2.7-58.9 cm H 2 O) and V T (1-14.6 mL) was observed in a sample of 120 volunteers during ventilation of a neonatal mannequin with a self-inflating bag.…”
Section: Discussionsupporting
confidence: 93%
“…However, some studies showed that the valve may be activated at an inconsistent, wide range of pressures, 27 frequently exceeding safe values. 25,27 In the present study, even though the pop-off valve was kept open, we observed peak pressures over 40 cm H 2 O (the activation pressure limit indicated by the manufacturer) in 52% of the cases. Kain et al 10 observed higher levels of PIP when the pop-off valve was kept closed; however, the authors tested different brands of resuscitators and found high variability in PIP and V T independently of the position of the valve (opened or closed).…”
Section: Discussionmentioning
confidence: 40%
See 1 more Smart Citation
“…encontró que el volumen corriente y la PIM entregadas por profesionales médicos era mayor cuando se usaban BAI, en comparación con dispositivos con pieza en T, sin importar las variables dependientes del operador individual. 16 En un segundo estudio, Roerhs y cols. mostraron que, tanto la capacitación del operador como la experiencia específica con el dispositivo, tenían un impacto significativo en la provisión de mayor PIM y volumen corriente si usaban la BAI; esta diferencia no se verificó cuando usaban el dispositivo con pieza en T. Los mismos autores hallaron una gran variación interoperador para todos los niveles de capacitación cuando se usaba la bolsa autoinflable.…”
Section: Baiunclassified
“…Unter brüsker Beatmung kann der applizierte Beatmungsdruck allerdings durchaus den vom Hersteller angegebenen Spitzendruck übersteigen [186,187]. Die Verwendung eines Systems mit T-Stück erlaubt im Gegensatz zu herkömm-lichen Beatmungsbeuteln eine bessere Kontrolle des applizierten Spitzendrucks und des Tidalvolumens, außerdem erleichtert es die Beatmung mit verlängerten Inspirationszeiten [187][188][189][190]. Allerdings ist die klinische Bedeutung dieser Beobachtungen noch nicht eindeutig klar.…”
Section: Hilfsmittel Zur Assistierten Beatmungunclassified