Low eddy current RF shielding enclosure designs for 3T MR applications

Abstract: We evaluated the RF shielding effectiveness and the gradient-induced eddy current artifacts of several shielding designs, and only the PBM showed positive outcomes for both aspects. Magn Reson Med 79:1745-1752, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
24
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
7

Relationship

3
4

Authors

Journals

citations
Cited by 34 publications
(24 citation statements)
references
References 50 publications
0
24
0
Order By: Relevance
“…As presented from earlier publications, the intense gradient field slew rate of EPI sequence induced eddy current on the solid copper shielding cage of the RF‐penetrable PET insert and suffered from the Nyquist ghosting artifact. To allow the PET ring operable with EPI, various designs and materials for the shielding cage with sufficiently high shielding effectiveness and low induced eddy current have been investigated . This alternate shielding configuration will allow simultaneous PET/MR studies during EPI sequence.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…As presented from earlier publications, the intense gradient field slew rate of EPI sequence induced eddy current on the solid copper shielding cage of the RF‐penetrable PET insert and suffered from the Nyquist ghosting artifact. To allow the PET ring operable with EPI, various designs and materials for the shielding cage with sufficiently high shielding effectiveness and low induced eddy current have been investigated . This alternate shielding configuration will allow simultaneous PET/MR studies during EPI sequence.…”
Section: Discussionmentioning
confidence: 99%
“…To allow the PET ring operable with EPI, various designs and materials for the shielding cage with sufficiently high shielding effectiveness and low induced eddy current have been investigated. 39 This alternate shielding configuration will allow simultaneous PET/ MR studies during EPI sequence.…”
Section: Discussionmentioning
confidence: 99%
“…This degradation is mainly due to the eddy currents induced in the Faraday cages when the gradient-intensive EPI sequence was applied, which heats up the detector module and the SiPMs, causing the gain of the SiPMs to drop and the dark counts to increase. A new Faraday cage design made with phosphor bronze mesh is expected to reduce eddy current generation while maintaining high shielding effectiveness [59]. We are also implementing active liquid cooling of the detectors to further mitigate effects of any remaining gradient heating.…”
Section: Discussionmentioning
confidence: 99%
“…To prevent coupling of the RF coil to the gradient coil, a mesh RF shield was used after consideration of different options . An optimal RF shield is necessary to both isolate the RF coil and also minimize the eddy current heating from the high SR, high gradient amplitude switching.…”
Section: Methodsmentioning
confidence: 99%
“…To prevent coupling of the RF coil to the gradient coil, a mesh RF shield was used after consideration of different options. 31,32 An optimal RF shield is necessary to both isolate the RF coil and also minimize the eddy current heating from the high SR, high gradient amplitude switching. The RF shield was mounted on a separate structure to allow easy removal with the RF coil, and to eliminate heat transfer from the shield to the gradient coil.…”
Section: Gradient Coil Designmentioning
confidence: 99%