Purpose
To evaluate the diagnostic accuracy of the CapitalBio
Mycobacterium
real-time polymerase chain reaction assay (CapitalBio test) testing of core needle biopsy (CNB) specimens for tuberculous lymphadenitis (TBL) and to compare it with Xpert MTB/RIF.
Methods
We retrospectively analyzed the medical data on patients with suspected peripheral TBL. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and area under the curve (AUC) of the CapitalBio test, Xpert MTB/RIF, and parallel test (positive result for either of these two tests) were calculated to evaluate their diagnostic efficacy compared with the final clinical diagnosis.
Results
The study included 114 patients. For diagnosing TBL using CNB samples, the sensitivity, specificity, PPV, NPV, and AUC were 65.0%, 100.0%, 100.0%, 28.6%, and 0.83, respectively, for the CapitalBio test; 72.0%, 100.0%, 100.0%, 33.3%, and 0.86, respectively, for Xpert MTB/RIF; and 82.0%, 100.0%, 100.0%, 43.8%, and 0.91, respectively, for the parallel test.
Conclusion
The accuracy of the CapitalBio test and Xpert MTB/RIF for diagnosing TBL using CNB specimens was moderate, while the sensitivity and NPV of these two tests were relatively low. The diagnostic accuracy of the CapitalBio test was slightly lower than that of Xpert MTB/RIF, but the difference between the two was not statistically significant. Parallel test might improve the diagnostic accuracy but not substantially over a single test.