Influence of infusion line compliance on drug delivery rate during acute line loop formation

Abstract: Infusion line compliance contributes to irregular drug delivery associated with vertical displacement of syringe pumps. Siphoning of the infusion line during patient care should be avoided, and flow rates of 1 ml/h or higher are recommended. Low-compliance infusion lines are indicated whenever highly short-acting vasoactive drugs at low delivery rates are administered.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
34
0
4

Year Published

2000
2000
2017
2017

Publication Types

Select...
5
1

Relationship

2
4

Authors

Journals

citations
Cited by 55 publications
(40 citation statements)
references
References 9 publications
2
34
0
4
Order By: Relevance
“…7 Infusion line compliance may further influence pump performance substantially. 9 The importance of using the original stiff, low compliant tube extension set with a very small inner diameter was demonstrated by the four-fold increase in ZDDT when it was replaced by the soft, more compliant Injectomat® infusion line (Table II). 9 The design of the tested PMV infusion device results in minimal mechanical gaps, a low overall system compliance and contributes to its optimized functional performance.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…7 Infusion line compliance may further influence pump performance substantially. 9 The importance of using the original stiff, low compliant tube extension set with a very small inner diameter was demonstrated by the four-fold increase in ZDDT when it was replaced by the soft, more compliant Injectomat® infusion line (Table II). 9 The design of the tested PMV infusion device results in minimal mechanical gaps, a low overall system compliance and contributes to its optimized functional performance.…”
Section: Discussionmentioning
confidence: 99%
“…9 The importance of using the original stiff, low compliant tube extension set with a very small inner diameter was demonstrated by the four-fold increase in ZDDT when it was replaced by the soft, more compliant Injectomat® infusion line (Table II). 9 The design of the tested PMV infusion device results in minimal mechanical gaps, a low overall system compliance and contributes to its optimized functional performance. 10 Even though this device may be very advantageous in terms of high flow accuracy and low susceptibility to vertical displacement several points with respect to its potential use in pediatric anesthesia and intensive care medicine must be mentioned: First, in case of accidental occlusion, the device's cut-out pressure is set at 4 bar, resulting in a maximum back-up bolus of 80 µL (manufacturer's specifications).…”
Section: Discussionmentioning
confidence: 99%
“…The compliance of the drug delivery tubing affects the variability of its infusion rate, affecting the time to initiate the occlusion alarm, time without flow, suction volume, and bolus volume during vertical movements of the syringe pump [16,17]. The impact of this factor is further enhanced in combination with high volume syringes (50 mL) and low flow programming.…”
Section: Factorsmentioning
confidence: 99%
“…Diese Schwankungen sind auf Veränderungen des hydrostatischen Drucks innerhalb des Infusionssystems zurückzuführen [6]. Das Ausmaß der Unregelmässigkeiten ist von der Compliance der Infusionsschläuche, der Infusionsspritzen und der Spritzenpumpen abhängig, die bei wechselnden hydrostatischen Durckverhältnissen beim Absenken der Spritzenpumpe zum Pooling von Infusat ins Infusionssystem und vice versa beim Anheben der Spritzenpumpe auf das Ausgangsniveau zum Entleeren eines Infusionsbolus in die venöse Strombahn führt [8,13].…”
Section: Zusammenfassungunclassified