2022
DOI: 10.1002/acm2.13835
|View full text |Cite
|
Sign up to set email alerts
|

Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT

Abstract: This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head‐and‐neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and each tissue‐equivalent plug of the “Cheese” phantom were obtained from TomoTherapy MVCT. Two IVDTs with different air CT numbers were created and applied to MVCT images of the HN anthropomorphic phantom and recalcula… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
2
0

Year Published

2022
2022
2023
2023

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 15 publications
0
2
0
Order By: Relevance
“…After correction, the beam data of the nominal robust and PTV plan were copied to each MVCT image, and the dose distributions were recalculated. Notably, in this study, three dose recalculation scenarios were considered because of the difference in the dose calculation accuracies of kVCT and MVCT (14,15). The dose for MVCT was approximately 2% higher than that for kVCT (14,17); therefore, the scenarios were dose correction of 1% (DC1%), DC2%, and DC3%.…”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…After correction, the beam data of the nominal robust and PTV plan were copied to each MVCT image, and the dose distributions were recalculated. Notably, in this study, three dose recalculation scenarios were considered because of the difference in the dose calculation accuracies of kVCT and MVCT (14,15). The dose for MVCT was approximately 2% higher than that for kVCT (14,17); therefore, the scenarios were dose correction of 1% (DC1%), DC2%, and DC3%.…”
Section: Methodsmentioning
confidence: 99%
“…Tomotherapy is equipped with a megavolt-CT (MVCT) system, and the acquisition of an MVCT image is a prerequisite before each treatment fraction (12). MVCT can be used for IGRT and dose recalculation (13)(14)(15). The present study aimed to compare the dosimetric consequences of robust and PTV-margin-based optimization to assess the efficacy of optimization technique on helical tomotherapy plans for prostate cancer in the presence of interfractional anatomical variations.…”
mentioning