Improper positioning of the elevator lever of duodenoscopes may lead to sequestered bacteria that survive disinfection by automated endoscope reprocessors

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
10
0
3

Year Published

2018
2018
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 15 publications
(13 citation statements)
references
References 8 publications
0
10
0
3
Order By: Relevance
“…As described by Alfa et al ( 20 , 21 ) BBF was formed over eight days at room temperature inside PTFE channels (Endoscopy Development Company, Maryland Heights, MO, USA) using ATS-2015 (Healthmark, Fraser, MI, USA) containing 8 Log 10 /mL (day 1) of E. faecalis and P. aeruginosa . On days 3, 4, and 5, the PTFE channels were rinsed and exposed to glutaraldehyde partial fixation (1:50 dilution of glutaraldehyde) and then repeat biofilm formation allowed to develop overnight ( 13 ).…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations
“…As described by Alfa et al ( 20 , 21 ) BBF was formed over eight days at room temperature inside PTFE channels (Endoscopy Development Company, Maryland Heights, MO, USA) using ATS-2015 (Healthmark, Fraser, MI, USA) containing 8 Log 10 /mL (day 1) of E. faecalis and P. aeruginosa . On days 3, 4, and 5, the PTFE channels were rinsed and exposed to glutaraldehyde partial fixation (1:50 dilution of glutaraldehyde) and then repeat biofilm formation allowed to develop overnight ( 13 ).…”
Section: Methodsmentioning
confidence: 99%
“…Once the BBF was fully formed on day 8, there was full HLD of the BBF using 2.6% glutaraldehyde (Metricide ® from Metrex—Sybron Canada, Oakville, ON, Canada) for 20 min at room temperature. Segments (5 cm) of the fully formed BBF were cut from the full PTFE-BBF channel and attached in between two 60 cm sterile PTFE segments to form a “surrogate endoscope channel” (SEC) that was 125 cm long as described by Alfa et al ( 20 , 21 ). The SEC model was used to mimic low levels of organisms within the BBF that was only present in the central 5 cm portion of the total instrument channel length.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…No breaches in adherence to manufacturer guidelines for high-level disinfection of scopes were identified in an outbreak investigation of carbapenem-resistant Klebsiella pneumoniae with blaOxa-232 gene associated with endoscopic retrograde cholangiopancreatography (ERCP) in 17 patients 19 . Reprocessing was less effective if the scope elevator mechanism was in a horizontal position as opposed to a vertical position during the high-level disinfection cycle in an automated endoscope reprocessor 20 . In yet another study, intraluminal fluid was detected in 22 out of 45 endoscopes tested after the completion of high-level disinfection.…”
Section: Reprocessing Of Endoscopesmentioning