2013
DOI: 10.1373/clinchem.2012.197020
|View full text |Cite
|
Sign up to set email alerts
|

Implications of Adjustment of High-Sensitivity Cardiac Troponin T Assay

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

1
7
0

Year Published

2013
2013
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 11 publications
(8 citation statements)
references
References 4 publications
(7 reference statements)
1
7
0
Order By: Relevance
“…Our findings corroborate and extend previous pilot studies 1113 and have important clinical implications beyond the specific hs-cTn assay examined. In our analysis, the hs-cTnT 2010–2012 low-end shift had no relevant clinical impact on the diagnosis of AMI when using conventional serial sampling, probably because the shift affected nearly exclusively values below the 99th percentile.…”
Section: Discussionsupporting
confidence: 90%
See 2 more Smart Citations
“…Our findings corroborate and extend previous pilot studies 1113 and have important clinical implications beyond the specific hs-cTn assay examined. In our analysis, the hs-cTnT 2010–2012 low-end shift had no relevant clinical impact on the diagnosis of AMI when using conventional serial sampling, probably because the shift affected nearly exclusively values below the 99th percentile.…”
Section: Discussionsupporting
confidence: 90%
“…22,[28][29][30] Our findings thereby confirm and reinforce previous expert statements that manufacturers need to ensure that their internal quality assurance programmes would detect a similar problem in the future. 11,13 As the novel early hs-cTn-based rule-out strategies are increasingly used clinically, 28 a similar low-end shift if occurring now would be expected to result in relevant errors in the diagnosis of AMI and thereby definitely harm patients. It is important that both manufacturers and clinicians take their lessons from the hs-cTnT 2010-2012 low-end shift.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…In addition Roche internally assessed the effect of affected lots with readjustment mathematically in 1033 patients presenting to the ER. 28 The identical number of MIs was defined.…”
Section: Limitationsmentioning
confidence: 99%
“…The limit of blank and limit of detection have been determined to be 3 ng/L and 5 ng/L, respectively (4 ). None of the hs-cTnT measurements in this analysis were affected by the 2010 to 2012 calibration shift (20,21 ). Estimated glomerular filtration rate was calculated using the abbreviated Modification of Diet in Renal Disease formula (22 ).…”
Section: Investigational Hs-ctn Measurementsmentioning
confidence: 99%