Impedance-guided Radiofrequency Ablation: Using Impedance to Improve Ablation Outcomes

Abstract: ABSTRACT. Despite the achievement of acute conduction block during catheter ablation, the recovery of conduction at previously ablated sites remains a primary factor implicated in arrhythmia recurrence after initial ablation. Real-time markers of adequate ablation lesion creation are needed to ensure durable ablation success. However, the assessment of acute lesion formation is challenging, and requires interpretation of surrogate markers of lesion creation that are frequently unreliable. Careful monitoring of… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

2
13
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 15 publications
(22 citation statements)
references
References 19 publications
2
13
0
Order By: Relevance
“…As described above, tissue heating by ablation results in a fall in GI 26,27 . Absent or suboptimal GI drop is therefore highly suggestive of ineffective current delivery 27 .…”
Section: Generator Impedance—tried and Trustedmentioning
confidence: 91%
See 1 more Smart Citation
“…As described above, tissue heating by ablation results in a fall in GI 26,27 . Absent or suboptimal GI drop is therefore highly suggestive of ineffective current delivery 27 .…”
Section: Generator Impedance—tried and Trustedmentioning
confidence: 91%
“…26,27 Absent or suboptimal GI drop is therefore highly suggestive of ineffective current delivery. 27 GI drops precede lesion maturation and may therefore result in premature termination of RF current. 28 After cessation of RF delivery, GI rises again, although it often remains below preablation levels.…”
Section: Generator Impedance-tried and Trustedmentioning
confidence: 99%
“…Thus, the absolute value and phase of bioimpedance are used to diagnose diseases, and this process is known as bioelectric impedance vector analysis [14][15][16][17][18][19]. Impedance measurement is also used in catheter ablation for treating arrhythmia [20][21][22][23]. By measuring the local impedance at an ablation site with electrodes integrated into an ablation catheter, the cauterization of myocardial tissue can be monitored.…”
Section: Introductionmentioning
confidence: 99%
“…The appropriate impedance decrease is critical for effective ablation. Excessive decrease in impedance will increase the incidence of complications, while a decrease in impedance that is too low indicates ineffective treatment [ 20 , 21 ]. The impedance changes rapidly when emergencies such as steam pops occur.…”
Section: Discussionmentioning
confidence: 99%