Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies

Abstract: The operation of the bowtie filter in x-ray CT is correct if the object being scanned is properly centered in the scanner's field-of-view. Otherwise, the dose delivered to the patient and image noise will deviate from optimal setting. We investigate the effect of miscentering on image noise and surface dose on three commercial CT scanners. Six cylindrical phantoms with different size and material were scanned on each scanner. The phantoms were positioned at 0, 2, 4 and 6 cm below the isocenter of the scanner's… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

7
95
2

Year Published

2014
2014
2024
2024

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 116 publications
(107 citation statements)
references
References 21 publications
7
95
2
Order By: Relevance
“…An exception was the eye lens, where the CT-Expo value of 60.6 mGy was higher than the measured mean value of 54.4 mGy. CT Expo values were lower however, for the breast (2.7 mGy vs 4.0 mGy) and thyroid Kaasalainen et al (7) and Habibzadeh et al (5) . These studies showed that when the scanning object was aligned off-center, below the isocenter, there was an increase in dose at the upper surface and an increase in image noise at the lower half of the phantom.…”
Section: Resultsmentioning
confidence: 99%
“…An exception was the eye lens, where the CT-Expo value of 60.6 mGy was higher than the measured mean value of 54.4 mGy. CT Expo values were lower however, for the breast (2.7 mGy vs 4.0 mGy) and thyroid Kaasalainen et al (7) and Habibzadeh et al (5) . These studies showed that when the scanning object was aligned off-center, below the isocenter, there was an increase in dose at the upper surface and an increase in image noise at the lower half of the phantom.…”
Section: Resultsmentioning
confidence: 99%
“…Optimal functioning of the filters requires appropriate centring of the patients. Miscentring can increase the surface radiation dose and image noise and introduce artefacts (17) . When two orthogonal localizers were acquired, a vertical miscentring of 5 cm had less effect on the tube current modulation (CTDI vol increased 1%) than using a single localizer.…”
Section: Discussionmentioning
confidence: 99%
“…From a clinical view point this is important as it has been shown that patient miscentering may lead to both higher surface doses and noisier images (10) . Positioning depends on the precision of the gantry lasers and the functionality of the tabletop of the CT scanner.…”
Section: Methodsmentioning
confidence: 99%