2005
DOI: 10.1002/elan.200503356
|View full text |Cite
|
Sign up to set email alerts
|

Highly Selective and Sensitive Copper(II) Membrane Sensors Based on 6-Methyl-4-(1-phenylmethylidene)amino-3-thioxo-1,2,4-triazin-5-one as a New Neutral Ionophore

Abstract: A new Cu(II) ion-selective PVC membrane sensor based on 6-methyl-4-(1-phenylmethylidene)amino-3-thioxo-1,2,4-triazin-5-one (MATTO) as an excellent sensing material was developed. The electrode exhibits a Nernstian slope of 29.2 AE 0.4 mV per decade over a very wide concentration range between 1.0 Â 10 À1 and 1.0 Â 10 À6 M, with a detection limit of 4.8 Â 10 À7 M (30.5 ng/mL). The sensor possesses the advantages of short conditioning time, fast response time (< 10 s), and especially, very good selectivity towar… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
14
0

Year Published

2006
2006
2010
2010

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 69 publications
(14 citation statements)
references
References 43 publications
(32 reference statements)
0
14
0
Order By: Relevance
“…Recently, several greatly selective and sensitive polyvinyl chloride (PVC) membrane ISEs for various metal ions have been reported [17][18][19][20][21][22][23][24][25][26]. In this work, we wish to introduce a highly Fe 3?…”
Section: Introductionmentioning
confidence: 98%
“…Recently, several greatly selective and sensitive polyvinyl chloride (PVC) membrane ISEs for various metal ions have been reported [17][18][19][20][21][22][23][24][25][26]. In this work, we wish to introduce a highly Fe 3?…”
Section: Introductionmentioning
confidence: 98%
“…Several highly selective and sensitive membrane sensors for alkaline earth and transition metal ions have been reported [8][9][10][11][12][13][14][15][16][17]. However, this study describes the development of a new ion selective sensor for the Ho(III) determination and, in particular, the use of 4-chloro-1,2-bis(2-pyridinecarboxamido)benzene (CBPB) as an ionophore for the preparation of a highly Ho(III) ion-selective electrode (Fig.…”
Section: Introductionmentioning
confidence: 99%
“…[12][13][14][15][16][17][18] Recently, several greatly selective and sensitive PVC-membrane ionselective electrodes for various metal ions have been reported. [19][20][21][22][23][24][25][26][27][28][29][30][31][32][33][34][35][36][37] Nevertheless, this paper focuses on the introduction of a highly siver(I)-selective sensor based on Dilaktam Crown ether (DLCE) (Fig. 1), as a novel neutral ionophore for monitoring silver concentration.…”
Section: Introductionmentioning
confidence: 99%