Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus

Abstract: Bias associated with the use of NaF-KOx tubes may have a significant impact on the prevalence of fasting hyperglycaemia, according to current diagnostic criteria. The small but significant difference between preanalytical processes should be considered when screening for the presence of diabetes mellitus.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
24
1
2

Year Published

2009
2009
2020
2020

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 39 publications
(27 citation statements)
references
References 14 publications
0
24
1
2
Order By: Relevance
“…The comparison of paired blood samples for serum & heparinized plasma collected and stored at same ambient temperature and centrifuged at same time produced higher glucose values in plasma [6]. On the other hand studies comparing the results of serum gel separator tubes with those of fluoride tubes have reported higher glucose values in serum samples [3,4,12]. This is because when serum is collected in tubes with a clot activator and serum gel separator, and promptly centrifuged, glycolysis is stopped quickly due to separation of serum from the cellular components [12].…”
Section: Dear Sirmentioning
confidence: 91%
See 1 more Smart Citation
“…The comparison of paired blood samples for serum & heparinized plasma collected and stored at same ambient temperature and centrifuged at same time produced higher glucose values in plasma [6]. On the other hand studies comparing the results of serum gel separator tubes with those of fluoride tubes have reported higher glucose values in serum samples [3,4,12]. This is because when serum is collected in tubes with a clot activator and serum gel separator, and promptly centrifuged, glycolysis is stopped quickly due to separation of serum from the cellular components [12].…”
Section: Dear Sirmentioning
confidence: 91%
“…Glucose levels can fall as much as 10 mg/dL during this period [2]. Transport on ice slurry and rapid separation of serum within 30 min can inhibit glycolysis without the addition of sodium fluoride and in fact works better as shown by few studies [3,4]. But this would be impractical to expect that every sample be cooled and separated soon after collection.…”
Section: Dear Sirmentioning
confidence: 99%
“…Serum insulin was measured in University Hospitals of Larissa using the Roche Modular system and the electrochemiluminescence immunoassay kit (Roche Diagnostics, UK). This assay showed .05 % cross-reactivity to intact human proinsulin and the primary circulating split form created by processing of the insulin prohormone [24]. Serum glucose concentrations were measured using an Olympus AU 2700 analyzer (Olympus UK, London, UK) incorporating an automated hexokinase method with ultraviolet detection at 340 nm.…”
Section: Methodsmentioning
confidence: 99%
“…Meticulous research is being done to minimize the use of multiple blood collection tubes without compromising the quality of glucose concentration in samples. 10,12 For measurements of blood levels of common clinical chemistry analytes like creatinine, lipid profile, bilirubin, liver enzymes etc, serum specimens are most favored. Serum separator gel tubes (SST) are being employed in clinical laboratories for testing most analytes.…”
Section: Introductionmentioning
confidence: 99%