Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
12
0

Year Published

2018
2018
2022
2022

Publication Types

Select...
5

Relationship

2
3

Authors

Journals

citations
Cited by 7 publications
(12 citation statements)
references
References 18 publications
0
12
0
Order By: Relevance
“…The gamma passing rates of the patient-specific verification demonstrated that the plans normalized by the transmission factor can be applied for clinical purposes when a clinically acceptable criterion (i.e., 2%/2 mm) is used [ 10 ]. For the SABR plan, the gamma passing rates of the original and normalized plans were poor with the 1%/1 mm criterion because SABR had small fields and high-gradient dose profiles.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The gamma passing rates of the patient-specific verification demonstrated that the plans normalized by the transmission factor can be applied for clinical purposes when a clinically acceptable criterion (i.e., 2%/2 mm) is used [ 10 ]. For the SABR plan, the gamma passing rates of the original and normalized plans were poor with the 1%/1 mm criterion because SABR had small fields and high-gradient dose profiles.…”
Section: Discussionmentioning
confidence: 99%
“…In recent decades, IMRT verification techniques have been developed to increase the efficiency and accuracy of associated QA procedures [ 10 ]. Plan verification is generally performed using detector array systems, ion chamber, or film before first treatment.…”
Section: Introductionmentioning
confidence: 99%
“…[10][11][12] PD has been established utilizing EPID for pretreatment verification. 13,14) 138 www.ksmp.or.kr Modern radiotherapy has increased the need for online dose verification owing to the complexity and accuracy of the new delivery techniques. (e.g., VMAT or Stereotactic ablative radiotherapy (SABR)).…”
Section: ) Portal Dosimetry (Pd)mentioning
confidence: 99%
“…IMRT modulates the photon-beam intensity by changing the position of the multileaf collimator (MLC) [10], thereby delivering a nonuniform fluence from any given position of the patient's treatment beam and optimizing the dose distribution [11]. How-ever, the calculation of the small or irregular fields frequently used in IMRT has been reported as inaccurate; even with state-of-theart dose calculation algorithms, the calculated dose distribution and the dose distribution delivered to the patient may differ [10]. Therefore, for all patients undergoing IMRT treatment, patient-specific quality assurance (QA) must be performed prior to radiotherapy [12].…”
mentioning
confidence: 99%
“…In particular, the gamma index analysis method, which compares and analyzes the calculated and measured doses, is the most used for patient-specific QA in IMRT treatment. In IMRT treatment, although the values of acceptable dose difference (DD) and distance-to-agreement (DTA) are not clearly defined, the clinically well-accepted values are 3% and 3 mm, respectively [10][11][12][13][14][15][16][17]. Our institution used these same values; as the passing criterion, such as DD or DTA increases, the passing gamma value will increase.…”
mentioning
confidence: 99%